Datex-Ohmeda S/5 Light Monitor and Light Solo
Table 2
Setting and resetting the error bits
Error bit Description
Setting the bit
Resetting the bit
BCOM Communication
between battery and
monitor is lost.
The bit is set if reading the test
byte from the battery is not
successful (test is done when
installing the battery pack into
the slot and during recharge
initialisation).
If test byte read is successful.
BDATA
Data in the battery
pack memory was
corrupted and the
battery memory has
been initialised by
factory defaults.
This bit will not be set.
If during recharge initialisation is
noticed that the battery has been
completely discharged and the
latest recharge has been completed
and been successful (slope detected
as a sign of completed recharge).
TEMPFLT Temperature
measurement of the
battery pack is
inoperative.
If the temperature measurement
result is not within specified
limits (poor contact, no contact
or short circuit on the battery
pack).
The check is done like in BCOM
and CELLFLT.
If battery temperature is within
limits:
-30
°
C < T < +70
°
C.
(Note in the Service View:
-300 < T < +700).
CELLFLT
Low voltage over the
battery pack.
Battery voltage is below 10 V.
The check is done like in BCOM
and TEMPFLT.
If battery voltage is over 10 V.
If error bits are set one at a time without any other bit being set (Table 1, bold text):
−
BCOM: either poor contact in the battery module or battery memory failure. Try in another
battery slot.
−
BDATA: Normally this does not appear alone. This will be reset after following procedure:
Recharge the battery to full, discharge completely empty and start recharging again.
−
TEMPFLT: either poor contact inside the battery module or a real temperature failure
(Temperature out of specifications). Try in another battery slot.
−
CELLFLT: either poor contact inside the battery module or failure on single/multiple battery
cell. Try in another battery slot. Monitor uses trickle charge for refreshing the battery. Fault is
cleared during recharge initialisation if voltage has arisen above 10 V.
Probably the most common error message is “Error 29” which means that all the possible failures
on the battery have occurred. This is commonly a problem with the contacts inside the battery
module. All the battery information is stored in the battery memory inside the battery pack and is
read upon main CPUs request. Contact failure can be checked in the Service view: if Status = FFh,
Faults = 1Dh, capacities are 24570 and charge/discharge cycles are 16383, it is clearly due to
poor contact.
One faulty battery pack can stock still the whole battery bus. In that case all the battery data is on
logical “0” or “1”. That is why the batteries should be checked one battery in the module at a time.
During Error 29 the battery window on the display is empty followed by a ? mark.
Document No. 890200-5
10
Summary of Contents for F-LM1
Page 4: ...Datex Ohmeda S 5 Light Monitor Document No M1037863 2 ...
Page 8: ...Datex Ohmeda S 5 Light Monitor Document No M1051339 iv ...
Page 68: ...Datex Ohmeda S 5 Light Monitor Document No M1051339 60 ...
Page 78: ......
Page 80: ...Datex Ohmeda S 5 Light Monitor Document No 8001934 2 ii ...
Page 88: ...Datex Ohmeda S 5 Light Monitor Document No 8001934 2 8 ...
Page 89: ...Appendix A Planned maintenance form APPENDIX A 9 Document No 8001934 2 ...
Page 90: ...Datex Ohmeda S 5 Light Monitor Document No 8001934 2 10 ...
Page 94: ......
Page 96: ...Datex Ohmeda S 5 Light Monitor Document No M1051339 ii ...
Page 112: ...Datex Ohmeda S 5 Light Monitor Document No M1051339 16 ...
Page 114: ......
Page 116: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 ii ...
Page 128: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 12 ...
Page 134: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 18 ...
Page 136: ......
Page 140: ...Datex Ohmeda S 5 Light Monitor 2 Document No 8000060 1 ...
Page 200: ...Datex Ohmeda S 5 Light Monitor 62 Document No 8000060 1 ...
Page 201: ...APPENDIX A Check form 63 Document No 8000060 1 APPENDIX A ...
Page 202: ...Datex Ohmeda S 5 Light Monitor 64 Document No 8000060 1 ...
Page 206: ......
Page 208: ......
Page 210: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 ii ...
Page 224: ......
Page 226: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 ii ...
Page 232: ......
Page 234: ...Datex Ohmeda S 5 Light Monitor Document No 890201 5 ii ...
Page 244: ......
Page 246: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No M1051340 ii ...
Page 254: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No M1051340 8 ...
Page 256: ......
Page 258: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No 890200 5 ii ...
Page 274: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No 896192 1 ii ...
Page 283: ...External Battery Charger N LCHGR 7 2 Exploded view Document No 896192 1 9 ...
Page 284: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No 896192 1 10 ...
Page 286: ......
Page 288: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 ii ...
Page 302: ...Datex Ohmeda Light Monitor Document No 8000060 1 14 ...
Page 304: ......
Page 306: ...Datex Ohmeda S 5 Light Monitor This page intentionally left blank Document No M1051358 ii ...
Page 316: ...Datex Ohmeda S 5 Light Monitor 65 66 64 Document No M1051358 10 ...
Page 325: ...Spare parts Document No M1051358 19 ...
Page 328: ...Datex Ohmeda S 5 Light Monitor This page intentionally left blank ...
Page 330: ......
Page 332: ...Datex Ohmeda S 5 Light Monitor Document No 896551 2 ii ...
Page 340: ...Datex Ohmeda S 5 Light Monitor Document No 896551 2 8 ...
Page 342: ......
Page 344: ...Datex Ohmeda S 5 Light Monitor Document No 896552 2 ii ...
Page 352: ......
Page 354: ...Datex Ohmeda S 5 Light Monitor Document No 896553 2 ii ...
Page 360: ......
Page 362: ...Datex Ohmeda S 5 Light Monitor Document No 896554 2 ii ...
Page 367: ...Method of Replacing the Lamp Unit LCD Backlight ...
Page 368: ......