Displays and Controller Boards
Document No. 8001004-1
35
5.1
Keyboard menu
The service menu for testing the command board functions and for setting the K-ARK and
K-ICU types.
Upper Led
is for testing the upper alarm LED (red) on the command board. When the text
is highlighted, the upper alarm LED can be turned on and off by pressing the ComWheel.
Lower Led
is for testing the lower alarm LED (yellow) on the command board. When the
text is highlighted, the lower alarm LED can be turned on and off by pressing the
ComWheel.
Dummy Press
is for testing the ComWheel. When the text is highlighted, pressing of the
ComWheel create a sound from the loudspeaker and the corresponding number on the
service data field increase.
Service Data
Message count
counts the number of messages that are sent out to the main CPU board.
Leds upper
and
lower
indicate the states of the alarm LEDs on the command board.
Direct action keys
texts are indications to the command board membrane keys. When a
key on the command board is pressed, the corresponding text in the menu changes its
colour.
Control wheel, Press
counts the ComWheel pressings.
Control wheel, Clockwise
and
Counterclockwise
the ComWheel turnings.
Since
shows the date and the time of the last run time reset.
Summary of Contents for D-VNC15
Page 2: ......
Page 6: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor Document No 8001004 1 iv...
Page 56: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor 50 Document No 8001004 1...
Page 57: ...Appendices A G Displays and Controller Boards Document No 8001004 1 51 APPENDICES A G...
Page 58: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor 52 Document No 8001004 1...
Page 60: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor A 2 Document No 8001004 1...
Page 62: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor B 2 Document No 8001004 1...
Page 64: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor C 2 Document No 8001004 1...
Page 66: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor D 2 Document No 8001004 1...
Page 68: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor E 2 Document No 8001004 1...
Page 70: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor F 2 Document No 8001004 1...
Page 72: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor G 2 Document No 8001004 1...