Displays and Controller Boards
Document No. 8001004-1
19
23.
Degaussing:
Select DEGAUSSING with the keys SELECT then push the key ADJUST (+) or (>). Check that
the picture on the screen sways briefly with a subdued sound
NOTE: Degaussing (demagnetisation) takes place automatically when the monitor is turned
on. The efficiency of a single degaussing phase decreases if repeated at intervals shorter
than 15 minutes.
?
24.
Color temperature adjustments:
Select COLOR or COLOR TEMPERATURE with the keys SELECT and ADJUST.
Check that the colors on the screen are changing with the keys ADJUST. Leave the
adjustment to a suitable level.
Perform the same steps also for RED, BLUE and GREEN (the last adjustment only in Rev.
00-02 displays).
?
25.
Contrast mode selection:
Select CONTRAST MODE.
Check that the contrast mode is turning on and off with the keys ADJUST. Leave the
selection into a suitable mode.
?
26.
Perform electrical safety check and leakage current test.
?
27.
Check that the display functions normally after the performed electrical safety check.
?
28.
Clean the picture tube with a clean, soft cloth dampened with suitable cleaning solution.
NOTE: Do not use any abrasives or hard objects to remove stains. Do not polish the screen.
Clean the display cover with a cloth dampened with a mild detergent solvent.
NOTE: Do not use strong solvents (e.g. acetone) or aerosol cleaners for cleaning the cover.
?
29.
Fill in all necessary documents.
Summary of Contents for D-VNC15
Page 2: ......
Page 6: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor Document No 8001004 1 iv...
Page 56: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor 50 Document No 8001004 1...
Page 57: ...Appendices A G Displays and Controller Boards Document No 8001004 1 51 APPENDICES A G...
Page 58: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor 52 Document No 8001004 1...
Page 60: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor A 2 Document No 8001004 1...
Page 62: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor B 2 Document No 8001004 1...
Page 64: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor C 2 Document No 8001004 1...
Page 66: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor D 2 Document No 8001004 1...
Page 68: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor E 2 Document No 8001004 1...
Page 70: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor F 2 Document No 8001004 1...
Page 72: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor G 2 Document No 8001004 1...