S/5 Anesthesia Monitor and S/5 Critical Care Monitor
18
Document No. 8001004-1
•
Prepare the picture for adjustments by pushing the menu key HELP on the Command Board
and removing one of the connected plug-in modules from the Central Unit. Wait until the
message regarding the removed module appears onto the monitor screen.
19.
Position adjustments:
Select VERT CENTRING with the keys SELECT and ADJUST.
Check that the vertical centring of the picture is changing with the keys ADJUST. Leave the
adjustment to a suitable level.
Perform the same steps also for HOR CENTRING.
?
20.
Size adjustments:
Select HEIGHT.
Check that the height of the picture is changing with the keys ADJUST. Leave the adjustment
to a suitable level.
Perform the same steps also for WIDTH.
?
21.
Shape adjustments:
Select PINCUSHION.
Check that the shape of the picture is changing with the keys ADJUST. Leave the adjustment
to a suitable level.
Perform the same steps also for TRAPEZOID and ORTHOGONALITY.
?
22.
Tilt adjustment:
Select TILT.
Check that tilting of the picture is changing with the keys ADJUST. Leave the adjustment to a
suitable level.
?
Summary of Contents for D-VNC15
Page 2: ......
Page 6: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor Document No 8001004 1 iv...
Page 56: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor 50 Document No 8001004 1...
Page 57: ...Appendices A G Displays and Controller Boards Document No 8001004 1 51 APPENDICES A G...
Page 58: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor 52 Document No 8001004 1...
Page 60: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor A 2 Document No 8001004 1...
Page 62: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor B 2 Document No 8001004 1...
Page 64: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor C 2 Document No 8001004 1...
Page 66: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor D 2 Document No 8001004 1...
Page 68: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor E 2 Document No 8001004 1...
Page 70: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor F 2 Document No 8001004 1...
Page 72: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor G 2 Document No 8001004 1...