S/5 Anesthesia Monitor and S/5 Critical Care Monitor
12
Document No. 8001004-1
Monitor ID register
The monitor ID register contains a three bit (numeric values 0-7) monitor ID code. The register is
connected to channels 11-13 of the X3 D-connector. If no display is connected to B-DISP board or
if the monitor ID code fails, 111 code is generated by pull-up resistors. If the ID code is 011 or
101, VGA resolution is selected, otherwise high resolution is selected.
Figure 3
Display Controller Board, B-DISP, block diagram
Summary of Contents for D-VNC15
Page 2: ......