![Datex-Ohmeda Cardiocap/5 Series User Manual Download Page 28](http://html1.mh-extra.com/html/datex-ohmeda/cardiocap-5-series/cardiocap-5-series_user-manual_3769770028.webp)
RECORDING AND PRINTING
You need
•
Built-in recorder for recordings
•
Thermal paper for the recorder
•
Laser printer for printouts
NOTE: Recordings on thermal paper may be destroyed when
exposed to light, heat, alcohol, etc. Take a photocopy for your
archives.
(1)
Record Waveform/Stop
key to start recording selected real-time
waveforms and to stop recording
(2)
Record Trend/Stop
key to start recording a numerical trend or
selected graphical trend and to stop recording
(3) Key to release the recorder paper compartment
(4) Recorder paper
(5) Recorder paper compartment
Recording waveforms
To record waveforms, press the
Record Waveform/Stop
key.
To stop recording, press
Record Waveform/Stop
again.
To configure the waveform recording:
1. Press the ComWheel to open
Main Menu
and select
Record/Print
.
2. Select
Record Waveforms
.
3. Select
Waveform 1
,
Waveform 2
or
Waveform 3
and choose the
parameter(s) for up to three waveforms.
4. Select
Delay
to choose when to start the recording:
(0, for recording only real-time data, or 12 seconds).
5. Select
Paper Speed
to adjust the paper speed (slower speeds
result in sharper images).
6. Select
Length
and choose the recording time (30 seconds or
continuous).
Recording on alarms
1. Press the ComWheel and select
Record/Print
.
2. Select
Record Waveforms
.
3. Select
Start on Alarms
and
YES
.
Recording is activated by Asystole, HR High/Low or P1
High/Low.
P1 and ECG waveforms are recorded. Selections are
preconfigured.
– 19 –