DT4240 xx xx xx xx
20230503-V0
55
Digital output test
Sequence of operations
8. Keep stepping by pressing
▲
button as long as you see
this menu item:
(
T
est:
D
igital
Out
put = digital
output test).
9. Enter the menu item by pressing the
ENTER
button.
10. You see:
on the display. Meaning L: limit
output (
L
imit), meaning the two underline: first and second
limit outputs are in off state.
11. Pressing the
▲
button, you can switch the outputs on or
off.:
•
: both are switched OFF.
•
: 1. ON; 2. OFF.
•
: 1. OFF; 2. ON.
•
: both are switched ON.
Rel1, Rel2 yellow indicators on the left side of the front
panel indicate the ON status of the limit outputs.
12. Press the
MENU
button, to exit from the menu item, and
you see this:
.
If you want finish test, continue from part
Exit from menu
item
.
Analog output test
Sequence of operations
13. Keep stepping by pressing
▲
button as long as you see
this menu item:
(
T
est:
A
nalog
Out
put = analog
output test).
14. Enter the menu item by pressing the
ENTER
button.
15. You see this:
(00 mA).
16. Pressing
▲
button you can increese the value of output
current in 1 mA steps in 0 - 20 mA range.
17. Press the
MENU
button, to exit from the menu item, and
you see this:
.
Summary of Contents for DT4240 Series
Page 1: ...DT4240 xx xx xx xx AC Voltage Meter Operating Instructions...
Page 68: ...DT4240 xx xx xx xx 68 20230503 V0 10 2 Application example Analog output option...
Page 69: ...DT4240 xx xx xx xx 20230503 V0 69 RS232 communication interface option...
Page 70: ...DT4240 xx xx xx xx 70 20230503 V0 RS484 communication interface option...
Page 71: ...DT4240 xx xx xx xx 20230503 V0 71 RS485 bus topology...
Page 74: ...DT4240 xx xx xx xx 74 20230503 V0...
Page 75: ...DT4240 xx xx xx xx 20230503 V0 75...
Page 76: ......