![Datcon DT4240 Series Operating Instructions Manual Download Page 43](http://html2.mh-extra.com/html/datcon/dt4240-series/dt4240-series_operating-instructions-manual_5128702043.webp)
DT4240 xx xx xx xx
20230503-V0
43
7.9. Limit output state display mode
(12. menu item)
Function
The state of limit outputs may present on the display. Here
you can set the condition of presenting the states. [Factory
default: dynamic]
Sequence of operations
1. Enter the menu by typing the user or supervisor code.
The way the code should be type in is described in Chapter
7.1.
You see this on the display:
.
2. Keep stepping by pressing the
▲
button as long as you
see this menu item:
(
L
imit
Di
splaying = limit
displaying).
3. Enter the menu item by pressing the
ENTER
button.
4. You see this on the display:
. The mnemonic is
blinking, signaling that it can be modify.
5. Pressing
▲
button select the displaying mode.
You can choose from the following:
•
(
Dynam
ic = dynamic)
It displays the limit output state as far as the
ENTER
button
is pressed. [Factory default]
•
(
Timed
= timed)
It displays the limit output state as far as the
ENTER
button
is pressed and 2.5 sec. long after releasing button.
•
(
Stati
c = static)
Pressing
ENTER
button the limit output display can be
switch on or switch off.
•
(
Chang
ing = changing)
Changing the state any of the limit outputs (switch on or
switch off) it displays the state automatically for 2.5 sec.
long.
Keeping the
ENTER
button pressed it displays the
measuring value.
•
(
Perio
dical = periodical)
It changes between the limit display and measuring display
mode in 2.5 sec. period.
Keeping the
ENTER
button pressed it displays the
measuring value.
Summary of Contents for DT4240 Series
Page 1: ...DT4240 xx xx xx xx AC Voltage Meter Operating Instructions...
Page 68: ...DT4240 xx xx xx xx 68 20230503 V0 10 2 Application example Analog output option...
Page 69: ...DT4240 xx xx xx xx 20230503 V0 69 RS232 communication interface option...
Page 70: ...DT4240 xx xx xx xx 70 20230503 V0 RS484 communication interface option...
Page 71: ...DT4240 xx xx xx xx 20230503 V0 71 RS485 bus topology...
Page 74: ...DT4240 xx xx xx xx 74 20230503 V0...
Page 75: ...DT4240 xx xx xx xx 20230503 V0 75...
Page 76: ......