1.
Note
Hinweis
Artificial Mains Networks/
L.I.S.N.s of this type may be
used with any AC or DC
voltage up to 250 V ac (3
phase 400 V), 50 Hz or 60 Hz
or DC voltages up to 400 V.
(options for higher voltages
available).
There is no operating voltage
adjustment, simply connect to
any source available or re-
quired.
There is one restriction: most
L.I.S.N.s can
not
be used with
115 V / 400 Hz because of 8
times higher ground current.
Also because of the high
ground currents due to the
CISPR 16 / VDE 0876 recom-
mendations, it is normally not
possible to use L.I.S.N.s on
power lines with ground current
safety switches (that cut off
power due to excessive ground
current.) Either a special power
line outlet without the ground
current switch must be in-
stalled (warning label required!),
or an isolating power line
transformer 1:1 must be used.
In any case, ground-connect
L.I.S.N.
before
connecting to
power line.
Netznachbildungen dieses Typs
können für Wechselspannungen
bis 250 Veff oder Gleichspan-
nungen bis 400 V dc eingesetzt
werden.
Wegen zu hoher kapazitiver
Ableitströme ist lediglich eine
Verwendung in 400-Hz-Netzen
nicht ohne weiteres möglich
(ggf. Rückfrage).
Gleichfalls wegen hoher kapa-
zitiver Ableitströme
(Bemessung der Ableitkon-
densatoren nach CISPR 16 /
VDE 0876) ist in der Regel ein
Betrieb mit Fehlstrom-
schutzschalter in der Gebäu-
deinstallation (bzw. im Meßla-
bor)
nicht
möglich.
Entweder muß eine Sonder-
steckdose ohne FI-Schalter und
entsprechende Beschriftung in-
stalliert werden, oder es wird
ein Netz-Trenntransformator
1:1 eingesetzt.
In jedem Fall beachten:
Jede Netznachbildung muß
vor
der Netzverbindung an Schutz-
erde gelegt werden.