i) Connector for Device under Test:
"Schuko"-plug for max. 16 A.
P. E. directly to ground
or via ground choke.
j) Reference Ground:
Bare ground
connector on front panel, milled
circular areas on the side panels,
k) R.F. output:
BNC-plug on the front
panel, 50
Ω
l) Cabinet:
19" Standard cabinet
17"x 5.2"x 16"
m) Safety:
Class 1 acc. to VDE 0100/
5.73/10.73 part 1 of VDE 0411,
chapter 8.10. P.E. via ground choke
acc. to VDE 0550, part 6 / 4.66
n) Ground current:
Acc. to VDE 0876,
part 1 (9.78), up to 1 A caused by
8
µ
F + 4
µ
F.
Warning on the front panels con-
cerning safety ground via a second
separate ground line.
i) Meßobjektanschluß:
220-V-Schuko-
dose, maximal 16 A.
Schutzleiter direkt an Masse oder
über Drossel für HF hochlegbar.
j) Meß-Bezugserde:
Vorn an Frontplatte
Erdungsschraubklemme mit 4 mm
Bohrung, an Rückseite 4 mm Telefon-
buchse, seitlich freigefräste Flächen.
k) Meßgeräteanschluß:
BNC-Buchse an
der Frontplatte, 50
Ω
l) Gehäuse:
Standardausf. im 19 Zoll -
Normgeh, 448x132x405 mm.
m) Schutzmaßnahmen:
Schutzklasse 1
nach VDE 0100/5.73/10.73 Teil 1
von VDE 0411, Abschn. 8.10.
Schutzleiter über SL-Drossel nach
VDE 0550, Teil 6 / 4.66
n) Ableitstrom:
Entsprechend der Be-
schaltung mit Ableitkondensatoren
nach VDE 0876, Teil 1 (9.78) 8
µ
F
+ 4
µ
F, bis zu 1 A.
Auf der Frontplatte befinden sich
Hinweise über Sicherstellung ein-
wandfreier Schutzerdung, ggf. über
zweite Leitung an separater Erde.