TECHNICAL SUPPORT
51
6
6.3.2 ERROR MESSAGES
ERROR MESSAGE
CAUSE
ACTIONS
S
DIO
DE
MA
X CU
FA
ULT
The laser diode max current
software level was exceeded. The
value is displayed in the Max.
Current field in the LASER DATA
section.
Turn off, wait 30/60 minutes and
restart. Read the value displayed in
the
Diode Current
field in the
LASER MONITOR section and
check to see if it exceeds the Max.
Current value. In case the problem
occurs, carry out the data log with
the service tool
and contact
technical assistance.
S
DIO
DE
HIG
H TE
M
FA
ULT
The laser diode max temperature
software level was exceeded. The
value is set at 35°C. The system
cuts off the current on the diode and
tries to bring the laser diode to the
proper temperature.
Turn off, wait 30/60 minutes and
restart. Make sure that the cooling
fans unit is working properly and
that the fans are not obstructed in
any way. Read the value displayed
in the
Diode Temp.
field in the
LASER MONITOR section and
check to see if it deviates from the
Set-point value. In case the
problem occurs, carry out the data
log with the service tool and contact
technical assistance.
S
DIO
DE
LO
W T
EM
FA
ULT
The laser diode temperature has
fallen below the minimum
temperature software level. The
value is set at 10°C. The system
turns off the entire power section
(laser diode, Peltier module, RF
driver, etc.) since the laser diode
temperature controller is out of
control.
Turn off, wait 30/60 minutes and
restart. Make sure the room
temperature complies with the
specifications. Read the value
displayed in the
Diode Temp.
field
in the LASER MONITOR section
and checks to see if it deviates
from the Set-point value. In case
the problem reoccurs, carry out the
data log with the service tool and
contact technical assistance.
S
DIO
DE
OUT
PUT
PRO
TEC
The laser diode voltage software
level was exceeded. The value is set
at 2.5V. The system turns off the
entire power section (laser diode,
Peltier module, RF driver, etc.) since
the laser diode may be broken, its
contacts not connected properly or
the internal card malfunctioning.
Turn off, wait 30/60 minutes and
restart. In case the problem re-
occurs, carry out the data log with
the service tool and contact
technical assistance.
S
TE
C
CURRENT
FA
ULT
The Peltier module current software
level was exceeded. The value is set
at 12A. The system turns off the
entire power section (laser diode,
Peltier module, RF driver, etc.) since
the Peltier module may be broken,
its contacts not connected properly
or the internal card malfunctioning.
Turn off, wait 30/60 minutes and
restart. Read the value displayed in
the
TEC Current
field in the
SYSTEM MONITOR section. In
case the problem reoccurs, carry
out the data log with the service
tool and contact technical
assistance.