Register Description
53
Note:
ACK (Acknowledge) and Read Back are not be available on the DT9841, DT9841E,
and DT9841-VIB module. To calibrate the module, the potentiometer must be reset by writing
the lowest value, and then incrementing it until the correct value is set.
After each byte of data is transferred, an ACK low-going signal can be read on the SDA line on
the 9th high-going, low SCL signal. The data is valid 4.7
μ
s after the SCL switches from low to
high and 0.3
μ
s after the high-to-low transition.
The first byte sent after a start command must be the device type followed by the address. The
device type is 0101 for a X9258; the addresses for the potentiometers are listed in
Table 7: Xicor Potentiometer Addresses and Functions
Address
Potentiometer
Function
0000
0
A/D 0 Zero Calibration
0000
1
A/D 1 Zero Calibration
0000
2
A/D 2 Zero Calibration
0000
3
A/D 3 Zero Calibration
0001
0
A/D 4 Zero Calibration
0001
1
A/D 5 Zero Calibration
0001
2
A/D 6 Zero Calibration
0001
3
A/D 7 Zero Calibration
0010
0
A/D 0 Full Scale Calibration
0010
1
A/D 1 Full Scale Calibration
0010
2
A/D 2 Full Scale Calibration
0010
3
A/D 3 Full Scale Calibration
0011
0
A/D 4 Full Scale Calibration
0011
1
A/D 5 Full Scale Calibration
0011
2
A/D 6 Full Scale Calibration
0011
3
A/D 7 Full Scale Calibration
0100
0
D/A 0 Zero Calibration
0100
1
D/A 1 Zero Calibration
0100
2
D/A 0 Full Scale Calibration
0100
3
D/A 1 Full Scale Calibration