Copyright © 2008 – 2019 DARE!! Products
Page 20
5.2
Commands Plug-In Card
The following table shows the general commands for the Radi
P
ower
®
plug-in card.
Command
Reply
Description
“*IDN?”
“USB1004A, Version
4.1.0”
Returns the model and version of
the device.
"ID_NUMBER?"
“d.d.d.d.d.d.d.d”
Returns the unique identifier
number of the Radi
P
ower
®
.
“POWER?”
“-12.34 dBm”
Get measure power level in dBm.
“BURST? <NUM>”
“-12.34 -12.35 dBm”
Get multi measure power level in
dBm.
The number of measurements is
requested in <NUM>.
"FREQUENCY?"
<NUM> Hz
Get the frequency in Hz.
"FREQUENCY <NUM> Hz"
“OK” or error code.
Set the frequency in Hz.
"FREQUENCY? MIN"
“9000 Hz”
Get the minimal frequency.
"FREQUENCY? MAX"
“6000000000 Hz”
Get the maximum frequency.
"FILTER?"
“1”, “2”, “3”, “4”, “5”,
“6”, “7” or “AUTO”
Get Filter.
"FILTER <NUM>"
“OK”
Set Filter to 1,2,3,4,5,6 or 7.
"FILTER AUTO"
“OK”
Set Filter to auto.
·
IEEE command for Interface Clear or Clear are not supported.
·
IEEE Status flags in either serial or parallel poll, or as a service request, are not supported.
·
When IEEE communication is used, the first command/request should be the "*IDN?\r"
command.