4. Installation
4.1 Wall mounted version (BH200LW)
The wall bracket needs to be true on the vertical plane.
4.1.1 Considerations
Wall Construction
Before fixing it is important to consider the strength of
the wall and the depth of the fixings used.
Use of a building engineer or structural drawings may be
required to decide on suitability.
Examinations to be
undertaken
We recommend the bracket be installed approximately
850-1100mm from the floor to the underside of the
bracket – dependent on couch height.
It may be necessary to adjust this height, if there are
specific medical procedures the light will be used for. If
in doubt consult the end user or DARAY. When installing
the bracket take consideration of where you want the
light head to sit when in its lowest position.
Locations of nearest
mains sockets
The BH200 wall mounted light uses a plug top transformer
with a DC jack connection.
Only the supplied certified power adapter should be used
with the light.
The length of this cable is approximately 1.8m so the wall
mounted light should be within 1.5m of a mains power
point