![Danmeter CSM 2 User Manual Download Page 21](http://html1.mh-extra.com/html/danmeter/csm-2/csm-2_user-manual_3418705021.webp)
Cerebral State Monitor - CSM DFU
Page 17
6 CSM Battery / CSM Power Operation
6.1
Using CSM Power and CSM Rechargeable Batteries
The CSM can be operated in two different modes
•
Battery operated
•
CSM Power operated
Warning:
Only standard CSM rechargeable battery CSMX04 or alkaline batteries must be used.
Battery operation (Alkaline batteries)
•
Always use new Alkaline batteries to ensure best results
•
Insert battery correctly
CSM Power operation and CSM rechargeable batteries
Warning:
Only for use with adapter CSMX05 (Ault Inc., Model MW117)
Note:
There is no alarm if the battery is not connected. If no battery is inserted, the monitor is switched off
when the external CSM Power is removed. There is no internal back-up battery in the monitor.
Note:
The clock is powered by an internal battery with a limited life (minimum 6 years from date of
manufacture). When the battery is used up, the clock will be reset, and the clock will blink on start
up. The monitor should then be sent for service.
Note:
To maintain and fully maximize the rechargeable battery lifetime, always connect the monitor to the
CSM Power when the device is not in use.
Battery Lifetime (Fully charged CSMX04 only at normal operation)
Wireless Link
Backlight level
Life time (hours)
OFF OFF 10.5
OFF 1 9.0
OFF 2 7.0
ON OFF 10.0
ON 1 8.5
ON 2 6.5
Note:
Using 9V alkaline disposable
batteries the battery life time
indicated in the table can be
multiplied by 2 to 3.
Summary of Contents for CSM 2
Page 2: ......
Page 33: ...Cerebral State Monitor CSM DFU Page 29 Notes...
Page 38: ......
Page 39: ......