Cerebral State Monitor - CSM DFU
Page 7
4.5
Operation Display Modes
The CSM always starts with display A
Switch between display A, B, C, D and E by pressing the Display key (7).
Display Mode D
CSI trend histogram with 5-minute interval showing average, lowest and highest
CSI values within the interval. EMG% is displayed as a bar in the right panel
Display Mode C
CSI trend curve with event markers and BS curve. EMG% is displayed as a bar in
the right panel. Operation time and numeric value of the actual CSI and BS%. The
time scale for the trend curve is 5.27 minutes and each tag is 60 seconds
Display Mode B
Displays 3 seconds of EEG waveform. Use Event key (8) to scale up and down. EEG
scales are ±200
μ
V, ±100
μ
V (default), ±50
μ
V, ±20
μ
V and ±10
μ
V
Display Mode A
Main information window with CSI, BS%, EMG%, SQI%, clock, operation time and
battery operation.
Display Mode E
.
Sensor impedance, CSI, BS%, clock and operation time. When display E is active,
the sensor impedance is updated every 60 seconds. Press Display key (7a) firmly
for immediate impedance update