Performing PRNU to a user entered value
Purpose:
Performs PRNU calibration to user entered value and eliminates the
difference in responsivity between the most and least sensitive pixel
creating a uniform response to light. Using this command, you must
provide a calibration target.
Executing these algorithms causes the
ssb
command to be set to 0
(no background subtraction) and the
ssg
command to 4096 (unity
digital gain). The pixel coefficients are disabled (
epc 0 0
) during
the algorithm execution but returned to the state they were prior to
command execution.
Syntax:
cpa i i
Syntax Elements:
i
PRNU calibration algorithm to use:
1
= This algorithm first adjusts each tap’s analog gain so that 8-
13% of pixels within a tap are above the value specified in the
target value parameter. PRNU calibration then occurs using the
peak pixel in the region of interest.
This algorithm is recommended for use only when FPN is
negligible and FPN coefficients are set to zero. Since this
algorithm adjusts the analog gain, it also affects FPN. If FPN is
calibrated prior to running this algorithm, FPN will be
observable in dark conditions and an incorrect FPN value will be
used during PRNU calibration resulting in incorrect PRNU
coefficients.
2
= Calculates the PRNU coefficients using the entered target
value as shown below:
PRNU
Coefficient
=
Target
(AVG
Pixel
Value
)
‐
(FPN
+
value)
sdo
i
i
i
The calculation is performed for all sensor pixels but warnings
are only applied to pixels in the region of interest. This
algorithm is useful for achieving uniform output across multiple
cameras. Is is important that the target value (set with the next
parameter) is set to be at least equal to the highest pixel across
all cameras so that all pixels can reach the highest pixel value
during calibration.
3
= This algorithm includes an analog gain adjustment prior to
PRNU calibration. Analog gain is first adjusted so that the peak
pixel value in tap’s ROI is within 97 to 99% of the specified
target value. It then calculates the PRNU coefficients using the
target value as shown below:
PRNU
Coefficient
=
Target
(AVG
Pixel
Value
)
‐
(FPN
+
value)
sdo
i
i
i
The calculation is performed for all sensor pixels but warnings
are only applied to pixels in the region of interest. This
algorithm is useful for achieving uniform output across multiple
cameras.
This algorithm is useful for achieving uniform output across
multiple cameras by first adjusting analog gain and then
performing PRNU calibration. This algorithm is recommended
for use only when FPN is negligible and FPN coefficients are set
to zero. Since this algorithm adjusts the analog gain, it also
affects FPN. If FPN is calibrated prior to running this algorithm,
Summary of Contents for S3-10-01k40-R
Page 13: ...Spyder 3 CL User s Manual 13 DALSA 03 032 20008 00 ...
Page 14: ...14 Spyder 3 CL User s Manual 03 032 20008 00 DALSA ...
Page 15: ...Spyder 3 CL User s Manual 15 DALSA 03 032 20008 00 ...
Page 16: ...16 Spyder 3 CL User s Manual 03 032 20008 00 DALSA ...
Page 24: ...24 Spyder 3 CL User s Manual 03 032 20008 00 DALSA ...
Page 44: ...44 Spyder 3 CL User s Manual 03 032 20008 00 DALSA ...
Page 102: ...102 Spyder 3 CL User s Manual 03 032 20008 00 DALSA ...
Page 110: ...110 Spyder 3 CL User s Manual 03 032 20008 00 DALSA ...
Page 112: ...112 Spyder 3 CL User s Manual 03 032 20008 00 DALSA ...