Linea ML Color Cameras
Appendix B: Trouble Shooting Guide
75
Image Quality Issues
Vertical Lines Appear in Image after Calibration
The purpose of flat field calibration is to compensate for the lens edge roll-off and imperfections in
the illumination profiles by creating a uniform response. When performing a flat field calibration,
the camera must be imaging a flat white target that is illuminated by the actual lighting used in the
application. Though the camera compensates for illumination imperfection, it will also compensate
for imperfections such as dust, scratches, paper grain, etc. in the white reference. Once the white
reference is removed and the camera images the material to be inspected, any white reference
imperfections will appear as vertical stripes in the image. If the white reference had imperfections
that caused dark features, there will be a bright vertical line during normal imaging. Similarly,
bright features will cause dark lines. It can be very difficult to achieve a perfectly uniform, defect-
free white reference. The following two approaches can help in minimizing the effects of white
reference defects:
1.
Move the white reference closer to or further away from the object plane such that it is out
of focus. This can be effective if the illumination profile changes minimally when relocating
the white reference.
2.
If the white reference must be located at the object plane, then move the white reference in
the scan direction or sideways when flat field calibration is being performed. The camera
averages several thousand lines when capturing calibration reference images so any small
imperfections are averaged out.
3.
Use the cameras flat field calibration filter feature, as detailed in the user manual Flat Field
Calibration Filter section. This algorithm implements a low pass moving average that covers
several adjacent pixels. This filter can help minimize the effects of minor imperfections in
the white reference. Note: this filter is NOT USED in normal imaging.
Over Time, Some Pixels Develop Low Response
When flat field calibration is performed with a white reference as per the guidelines in the user
manual, all pixels should achieve the same response. However, over time, dust within the lens
extension tube may migrate to the sensor surface thereby reducing the response of certain pixels.
If the dust particles are very small, they may have only a minor effect on responsivity, but still
produce vertical dark lines that interfere with defect detection and need to be corrected.
Repeating the flat field calibration with a white reference may not be practical with the camera
installed in the system. The camera has a feature where the flat field coefficients can be
downloaded to the host PC and adjusted using a suitable application, such as Microsoft Excel. (See
section Setting Custom Flat Field Coefficients for details.) If the pixel location that has a low
response can be identified from the image, then the correction coefficient of that pixel can be
adjusted, saved as a new file, and then uploaded to the camera, thereby correcting the image
without performing flat field calibration.
See the user manual for details on downloading and uploading camera files using CamExpert.
Note that dust accumulation on the lens will not cause vertical lines. However, heavy accumulation
of dust on the lens will eventually degrade camera responsivity and focus quality.
Summary of Contents for ML Series
Page 16: ...14 The Linea ML Color Camera Linea ML Color Cameras Mechanical Drawings ML FC 08K10T 00 R ...
Page 17: ...Linea ML Color Cameras The Linea ML Color Camera 15 ML FC 16K04T 00 R ...
Page 18: ...16 The Linea ML Color Camera Linea ML Color Cameras ML HC 16K10T 00 R ...
Page 80: ...78 Declaration of Conformity Linea ML Color Cameras Declaration of Conformity ...