54
Appendix A: GenICam Commands
Linea HS 32k CMOS TDI Cameras
Flat Field Category
The Flat Field controls, as shown by CamExpert, group parameters used to control the FPN and
PRNU calibration process.
Figure 23: Flat Field Panel
Flat Field Control Feature Description
Display Name
Feature
Description
View
Flat Field Correction Mode
flatfieldCorrectionMode
Beginner
DFNC
Off
Off FPN and PRNU correction disabled.
On
On FPN and PRNU correction enabled.
Clear Coefficents
flatfieldCalibrationClearCoefficient
Reset all FPN to 0 and all PRNU
coefficients to 1.
Beginner
DFNC
Calibration Algorithm
flatfieldCorrectionAlgorithm
Selection between four different PRNU
algorithms.
Beginner
DFNC
Peak
Peak Calculation of PRNU coefficients to bring
all pixels to the peak.
Peak, Image Filtered
PeakFilter A low pass filter is applied to the average
line values before calculating the
coefficients. Use this algorithm if the
calibration target is not uniformly white
or if it is not possible to defocus the
image. Because of the low pass filter, this
algorithm is not able to correct pixel-to-
pixel variations and so it is preferable to
use the “Peak” algorithm.