ZX-5
Ultrasonic Thickness Gauge
39
Note:
Once the measurement screen is displayed the
and
can be
released.
2) Press the
keys to scroll through the factory setting options.
3) Make a note of the “MEDI” & “ZERO” settings prior to performing a reset.
These values will need to be entered back in the gauge following the reset.
4) Press the
keys to scroll “REST” (reset).
5) Press the
key to edit the reset option.
6) Press the
keys to toggle YES, followed by pressing
to reset the gauge.
7) Repeat the steps above to set “MEDI” & “ZERO” back to their original
settings noted in step three above.
Summary of Contents for Z-303-0001