<Fig.3.3.22> Repeat Mode OFF
<Fig.3.3.23> Repeat Mode ON
(3) Hold button: It temporarily stops while in operation.
⑥
Repeat
-
You can set to repeat the operation within the selected Pattern by entering Repeat after
selecting Pattern from Program Settings screen.
-
If Pattern is created from Program Settings, you can set and store Repeat Mode regardless
of Operation Mode. However, when operating with Program Mode, you must set Repeat
Mode to ON.
-
(1) You set the number of repetition after verifying the Pattern number of Select Pattern.
You can set to repeat up to 100 times, and can be set to
ㆀ
infinity by pressing – button at
1.
-
(2) You can edit the order of operation after setting the number of repeat. Select Start
Segment and End Segment by pr button from Sequence list, and set the number
of repetition between the Sequences selected.
(A) Start Segment: Select first Segment for repeat operation
(B) End Segment: Select final Segment for repeat operation
(C) Segment Count: Set the number of repeats from Start Segment till End Segment
⑥
(1)
(A)
(B)
(C)
(2)