DAIHAN Scientific, Smart-Lab Temp./ Humidity Chamber 24/66
<Fig.3.3.4> Operation Settings
-
It is described in order for the operation of product. Proceed with the following steps in
order.
-
In order to use the product’s WiRe
TM
Service, please refer to how to install "WiRe
TM
Service".
①
Operation Settings
(1)
Running Mode After Power Failure
-
It sets a mode for power is abruptly cut off and turned back up due to power supply
issues such as power outage.
-
When setting a Stop mode: It stands ready at the value prior to the power outage and
maintains the operation on hold.
-
When setting a Restart mode: It operates depending on (4) Operation Mode.
A) Fixed Mode: When reauthorizing the power, it automatically operates at a temperature
set that it previously operated on and the operation time on main screen starts from
0 again.
B) Program Mode: It restarts from the first segment of the previous pattern it operated on.
-
Setting a Continue mode: It operates depending on (4) Operation Mode.
A) Fixed Mode: When reauthorizing the power, it automatically operates at a temperature
set that it previously operated on and the operation time on main screen
continues its counting.
B) Program Mode: It restarts the count from the remaining time of first segment that
stopped and it restarts.
(2)
Chart Display Cycle
-
You can set the temperature cycle shown on Chart. User can select from 30 seconds, 1
minute and 2 minutes.
(1)
①
(4)