DAIHAN Scientific, Smart-Lab Temp./ Humidity Chamber 22/66
3. Product Usage
3.3 Usage Instructions
-
If turning ON the power of this product, please turn ON the Main Power Switch after turning
ON the Leakage Breaker located at the bottom of the rear panel.
-
On the contrary, if turning OFF the power of this product, please turn OFF the Leakage
Breaker after turning OFF the Main Power Switch.
-
Prior to perform the operation of the product, the SAFETY Knob should be set up about
20~30
higher
℃
than service temperature.
-
Please operate after at least one minute when you turn the power ON again after turning
the power OFF.
-
Please check that the internal shelf is properly installed before the first operation.
-
Please set to OFF of Drain Button Before the first operation.
3.3.1. TEMPERATURE setting and operation
-
You can set the temperature by entering the number directly on the Temperature Setup
screen by pressing “SET” button on the right where the set temperature is shown.
-
ThermoStable STH-E Model can set the temperature at a 0.1
℃
increment and can set
between -20
℃
to 80
℃
. Press “ENTER” after entering the set temperature. It returns to the
main screen when the set temperature is applied.
<Fig.3.3.1>
Operation temperature setting