Be  careful  for  electrical  shock.  The  inappropriate  power  connection  induces  the 
damage of product or the injury of the human body.

This symbol indicates Protective earth (ground).

This symbol indicates power “OFF” on the power on/off switch.

This symbol indicates power “ON” on the power on/off switch.

This symbol indicates the connection point of the neutral conductor on the installed 
unit.

This symbol indicates alternating current Voltage.

1. General                                                                                                                       

      1.2 Product introduction

The Dry Block is compact and based on the ergonomic design to be conveniently applicable to 
narrow  spaces.  The  digital  control  employed  therein  can  realize  uniform  reproducibility  over 
entire  interval  of broad  range of temperature. The  product thus  can  be  applicable to  various 
labs in diverse disciplines and has the following features.

1. The Digital Feedback Control, controlled by the embedded microprocessor, enables users to 

have precision control and keep an accurate temperature. 

2. The high-quality insulation used in the product can prevent overheating and will keep the 

surface temperature of the main unit to a minimal level. 

3. The  high-performance  heater  integrated  into  the  product  was  designed  to  minimize 

defective rates through effective thermal conductivity. 

4. The  aluminum  Block  with  excellent  thermal  conductivity  reduces  the  heating  time,  and 

provides effective temperature distribution.

5. The wide display was furnished for the product to enable users a convenient checking on 

current range of set values for the operation of the product. 

6. The touch buttons were made by an excellent design focused on the convenience of users 

and maintenance thereof. 

Summary of Contents for MaXtable H20

Page 1: ...ed Validation IQ OQ Service Patented Patent No 10 0554098 Jog Shuttle Control System User Manual Version 2 2 1 written for MaXtableTM H20 MaXtableTM H20 High Performance Dry Bath Heating Block DAIHAN...

Page 2: ...Use 3 1 2 Product Introduction 4 1 3 Product Configuration 5 2 Installation 2 1 Product Installation 6 2 2 Product Specifications 7 3 Product Usage 3 1 Precautions for Use 10 3 2 Names of Each Part 1...

Page 3: ...he meaning Warning If the product is mishandled while ignoring this symbol the danger by which human beings are dead or severely injured may occur Caution Failure to follow the instructions may result...

Page 4: ...r entire interval of broad range of temperature The product thus can be applicable to various labs in diverse disciplines and has the following features 1 The Digital Feedback Control controlled by th...

Page 5: ...product is furnished with functions capable of protecting the overheating overcurrent and sensor errors of the product to secure the safety of users 9 The product is furnished with various kinds of b...

Page 6: ...together with the product is damaged and connect it with the product 4 Insert the socket in the other side into the wall plug In case of using a power strip be careful not to exceed the electrical en...

Page 7: ...uct Specifications Models MaXtableTM H20 Only Unit MaXtable TM H20 set included Standard Block Controller Single Digital Fuzzy Controller Useful Block Interchangeable Block 90 120 h60mm for Various Tu...

Page 8: ...t may cause overheating or electrical fault of the product This product was manufactured to move normally at the rated voltage Confirm the voltage state before the installation of product AC 230V 50 6...

Page 9: ...ve sufficient electric capacity The fluctuation of main power voltage should be lower than 10 of the nominal voltage Please separate the power plug from the outlet for safety when the product is unuse...

Page 10: ...the product to other place Please wear the protective equipment because the heating temperature can be overheated above the setting temperature Please always use the safety equipment working clothes...

Page 11: ...ion Power Switch Equipment power ON OFF Fuse Holder Fixing the fuse and protection not to expose Specification sticker Recording the specification and serial number of product Safety symbol sticker De...

Page 12: ...on Mode change button Timer setting mode Temperature setting mode Locking mode LCD Display Display operation information Rotary Switch Turn the clockwise or counterclockwise to adjust the setting valu...

Page 13: ...The range of values of the time set by the timer can be up to 99 time and 59 minutes and the set value will be varying by a unit of 1 minutes 6 Push the Rotary Switch once and shortly then the product...

Page 14: ...se to adjust the set value Press the PUSH button to have the set value fixed on the display The Reservation Timer for reserved operation of product will then begin 4 On completion of the pre setting o...

Page 15: ...value of current temperature whereas the figures displaying on the SV are the values of set temperature 3 Rotate the rotary switch on the controller clockwise or counterclockwise to adjust the set va...

Page 16: ...eleasing the MODE button 3 The scope of the compensative value is 10 0 10 0 and it is possible to adjust at 0 1 unit 4 After setting the compensation value the indication of the current temperature is...

Page 17: ...ll be changed into the display of locking mode 2 Accelerating and stop right after the locked display will prevent potential risks of malfunctioning of the product due to unintended turning of Rotary...

Page 18: ...Centrifuge Tube 48 holes DH WHB521248 Block BLT248 for 12mm Test Tube 48 holes DH WHB521335 Block BLT335 for 13mm Test Tube 30 holes DH WHB521535 Block BLT535 for 15mm Test Tube 30 holes DH WHB521628...

Page 19: ...y using driver Identify the fuse and replace it 120V F5AL 250V 230V F3AL 250V Re install the fuse holder to the body while pushing the holder with hands Classifications A period of checking time Day W...

Page 20: ...sed wipe off only the part with the towel wetting with the solvent that has low freezing point 6 If the materials wetted by liquid or water are entered into this product please immediately disconnect...

Page 21: ...t Switch off of the residual current device at the installation place or occurrence of power failure Inaccurate connection between a power plug and a socket Damage of socket plug power cable Disorder...

Page 22: ...cess about all functions before shipment 2 This instrument is warranted for 2 years from purchasing date 3 A breakdown caused by customer s improper use or natural disaster is excepted from this warra...