![Daihan Scientific MaXtable H20 User Manual Download Page 16](http://html1.mh-extra.com/html/daihan-scientific/maxtable-h20/maxtable-h20_user-manual_3214313016.webp)
DAIHAN-Sci.com or
ALL
for
LAB.com
Shopping Mall
DAIHAN Scientific, High Performance Dry Bath Heating Block "MaXtable
TM
H20"
16/22
3.3.4 Setting temperature Offset
1. Offset value is the function that can add or reduce the
constant value by setting the compensation value in the
case that user must experiment by matching the
thermometer which he/she is using as the experimental
standard and the thermometer indicated in this regulator.
2. Offset setting screen can be changed by holding down the
MODE button on the temperature setting screen and briefly
pressing the rotary switch once, and releasing the MODE
button.
3. The scope of the compensative value is -10.0℃ ~ +10.0℃,
and it is possible to adjust at 0.1℃ unit.
4. After setting the compensation value, the indication of the
current temperature is displayed as the temperature
including the compensation value.
5. If you enter the mode of Setup of Correction Values during
heating of the product, the heating of the product will stop.
Resume the heating of the product from the display of
setting temperature values after completing the Setup of
Correction Values.
6. Push the rotary switch for 2~3 seconds then the set value
will be blinking and be stored thereby. The stored value that
remaining unchanged will be reappearing by the next turn-
on of power supply.
7. If pressing Mode button, it is switched to the Temperature
setting screen.