Table 2
For all ME EQUIPMENT and ME SYSTEMS
IGuidance and manufacturer’s declaration - electromagnetic
immunity
The PM-150V is intended for use in the electromagnetic environ-
ment specified below. The customer or the user of the PM-150V
should assure that it is used in such an environment.
Prueba de
emisiones
Nivel prueba
IEC 60601
Nivel de
conformidad
Ambiente
electromagnético -
instrucciones
Electrostatic
discharge
(ESD)
IEC 61000-
4-2
± 6 kV
contact
± 8 kV air
± 6 kV
contact
± 8 kV air
Floors should be
wood, concrete
or ceramic tile. If
floors are covered
with synthetic ma-
terial, the relative
humidity should be
at least 30 %.
Power fre-
quency
(50/60 Hz)
magnetic field
IEC 61000-
4-8
3 A/m
3 A/m
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial
or hospital environ-
ment.