26
product and take it to an authorised Technical Support Service in the
event of any damage . Never use spare parts from any other brand.
No component can be maintained by user in the monitor. The circuit
diagrams, component part lists, descriptions, calibration instructions,
or other information which will assist the user’s appropriately qualified
technical personnel to repair those parts of equipment which are
designated repairably can be supplied.
It is recommended the performance should be checked every 2 years
or after repair. Please contact the service center.
The monitor can maintain the safety and performance characteristics
for a minimum of 10,000 measurements or three years
The cuff integrity is maintained after 1,000 open–close cycles of the
closure.
TECHNICAL DATA
Measurement method
Oscillometric. This means
that the product detects the
movement of blood through the
brachial artery, transforming
the blood pressure in a digital
readout.
Measuring Range
Cuff pressure: 0-300mmHg
Systolic: 60-260mmHg
Diastolic: 40-199mmHg
Pulse rate: 40-180 beats/
minute
Precision of the indication
±3 mmHg Pressure
Pulse ± 5% of indicated value
Memories
4x30
Operating Conditions
From +10 ºC to 40 ºC
Less than 85%
Transport and storage conditions
From -20 ºC to 50 °C
Less than 85%
Power
4 X 1.5V SIZE AAA batteries (not
included)
Battery Life
Approx. 100 times