15
F2. Calibrate the device: Contact an authorized technical service in case the
product shows abnormal temperatures.
F3. Activate/deactivate sound: Once inside the settings menu, when “F3” ap-
pears on the screen the device will be ready for you to activate or deactivate
the sound from the device. By pressing the up key (6) or down key (9) the
sound will be activated or deactivated.
F4. Changing modes: Once inside the settings menu, when “F4” appears on
the screen the device will be ready for you to calibrate. By pressing the up
key (6) or down key (9) the device will change between body temperature or
surface/object (SurF).
Changing measurement unit: Press and hold the measurement mode key for
some seconds to change from ºC a ºF and viceversa.
DATOS TÉCNICOS
Range of reading
Body mode: 32.0ºC – 42.9ºC (89.6ºF –
109.2ºF)
Surface mode: 0.0ºC – 100.0ºC (32.0ºF
– 212.0ºC)
Precision
Body mode: ±0.2ºC (±0.4ºF)
Surface mode: ±1ºC (±1.8ºF)
Screen resolution
0,1ºC/ºF
Memories
32
Operating Conditions
From 10ºC to 40°C
Humidity less than 80%
Storage and transport conditions
From -25ºC to 55°C
Humidity less than 90%
Power
2 AAA batteries (LR03)
Auto-power off
30 s
Time of reading
0,5
s
This product complies with the European Directive of medical product 93/42/EEC.