5) AUTO ZERO mode
Initialize sensor and indicator.
Execute this mode at unlock mode.
Execution process (display : operation mode)
(1) Sensor : no load.
(2) Voluntary data display.
(3) Press
key and
key at the same time.
Zero display (ex : 0, 0.0, 0.00, 0.000)
(4) This data is memorized to indicator though power turn off.
(5) If operate mode with input signal to remain, this point is memorized zero point.
6) AUTO CALIBRATION mode
Convenience to use when do not know calibration data of sensor or
displayed the measurement value differently.
(1) Adjust zero point of indicator.
(2) Apply known load to sensor. Large load is high safety.
(3) Press
key: F.S and CAL L.E.D turn on at the same time.
(4) As example table, 100.05 or 99.05 will blink.
(5) After correction of 100.00 using
key and
key, press
key.
(6) 100.05 is display correction of 100.00.
(7) At this time, cal data is corrected minutely.
(8) If corrected cal data is over limits of 2-4), correct of DIP S/W setting and again
set the auto calibration.
EX) At sensor Supply : 100.00 kg/
㎠
Display : 100.05 or 99.05 kg/
㎠
Adjustment : 100.00 kg/
㎠