3) PROGRAM (SOFTWARE) OPTION
(1) Press mode key, DP
+
CAL LED turn on at the same time.
mode
display
description
0 x x x
peak hold on : display maximum measurement data
peak hold off : display present measurement data..
reset : stored maximum measurement data is removed.
Adjustment of peak
hold mode
1 x x x
peak hold on(only) : display and reform
maximum measurement data
peak hold off : stored maximummeasurement data is
removed automatically.
mode
display
description
x 0 x x
standard type.
ex)
+
20mA : display 1000.0
Parity of
Display
x 1 x x
display (
+
) to (
-
)
signal and display indicate opposite.
ex)
+
20mA : display -760.0
x x 0 x
standard type.
Display 0 : trace minute signal to 0
Tracking
mode of
zero point
x x 1 x
Display 0 : display minute signal.
Use in case of leakage measurement.
x x x 0
display measurement signal immediately.
Average
value
display of
Signal
x x x 1
display average value of measurement
signal (7 times sampling).
7. ALARM OUTPUT (RELAY OUTPUT)
High and low relay are a and b contact respectively.
Contact capacity is maximum AC 250V/5A and DC 30V/5A.
Relay contact operating time is maximum (1/Sampling Time) mm Sec
* Output of transistor NPN open collector is option.
mode
setting
relay
high data > setting value
ON
HIGH
RELAY
high data
≤
setting value
OFF
low data
≤
setting value
ON
LOW
RELAY
low data > setting value
OFF