49
Sterilization time minimum 30 minutes at 121 °C or minimum 3 minutes at 135 °C,
and equivalent temperature and time parameters may be used within this range.
Maximum sterilization temperature 135 °C.
Step 4:
Sterilize the wrapped tip in a steam autoclave.
Step 5:
Make sure the drying cycle is complete before removing the tip from the autoclave. If the
bag is damp, proper sterilization cannot be guaranteed.
Always autoclave the tip wrapped in a sealed sterilization bag; failure to do
so will result in permanent smudges on the mirror.
Do not autoclave the device handpiece.
Tips should not be placed in an ultrasonic cleaner or any cold sterile solution.
Sterilizing solutions will leave a sticky residue or film on the mirror during
drying.
Do not remove the bag until the autoclave has completed its full drying cycle.
If the bag is wet or shows any signs of moisture, it can leave water spots on
the mirror, which can affect the image quality when scanning.
Be very careful when cleaning the mirror as it is very delicate and prone to
scratches.
Summary of Contents for eagle.IOS
Page 1: ......
Page 2: ...2 THIS PAGE IS INTENTIONALLY BLANK...
Page 25: ...25 A B System C D...
Page 69: ...69...
Page 70: ...70...
Page 71: ...71...
Page 72: ...72...