52
ITEM
DESCRIPTION OF THE INSPECTION
RECOMMENDED
FREQUENCY
Operation / Security
System
Activation of the on / off button (Visual).
Daily
Electrical parts
Overheating/Noise/Burn Smell (Auditory and
Visual).
Monthly
Parts and pieces
Operation/Noise/Vibration/Breaks (Auditory and
Visual).
Monthly
If problems are found during the inspection, contact the Alliage Service Department.
In addition to the annual inspection, to ensure a long durability and proper operation of your
equipment, it is important to perform preventive maintenance in a maximum period of three (3) years.
Contact the Alliage Service Department about our periodic review and preventive maintenance
program.
7.3.PREVENTIVE MAINTENANCE
7.4.CORRECTIVE MAINTENANCE
Do not open the equipment or try to fix it yourself or with the help of someone
without training or authorization. This can aggravate the problem or produce
a failure that may compromise the safety of the equipment.
Power cables and electronic boards can be changed only by the authorized
technician.
The equipment or any of its parts may not be repaired, maintained, or assisted
while being used on a patient.