background image

22 

 

displaying the ‘Last Time’ and ‘Next Time’, enter the number of days, then save in the top right 
corner. For accurate testing it is suggested for this to be set to 30-45 days, after this time has 
elapsed a warning will appear on the screen in the app to remind you that calibration is due. 
 
To exit tap the blue arrow in the top left. 
 
The end of the pH probe should not be allowed to dry fully. If the probe is not in use or being stored 
the protective cap should be replaced with enough probe storage solution in it to cover the end. If 
the probe is removed from the machine during maintenance or whilst calibration of other 
components is taking place it can be carefully stood in a small amount of aquarium water or the 
probe cap replaced as above. Never use pure water (RODI) to store the pH probe. 
 
 

CALIBRATING THE SALTWATER PUMP (TANK WATER IN) 

 
Go to the calibration settings in the Kamoer Remote app and select ‘Saltwater Pump’ from the 
common list. The ‘Saltwater Pump Calibration’ page will show you details of when the pump head 
was last calibrated, when the next calibration is due (this is a user defined setting), the current head 
calibration data and the option to ‘Go to calibrate’. 
 

 

To begin tap on ‘Go to calibrate’ in the app. 

 

Make sure you have the glass beaker and digital scales ready to perform the calibration, 
remove the front cover from the KH Manager by pulling it forwards exposing the tubes. 

 

When prepared tap ‘I’m ready’ in the app. 

 

Place the glass beaker in the base of the unit under the short tube that is on the far right as 
you look at the unit. 

 

In the Kamoer Remote app tap on the ‘Run’ button to start the pump head and flush the 
tubing. Once water is flowing into the glass beaker and there are no air bubbles tap the 
‘Stop’ button. 

 

In the app add a run time for the pump calibration by tapping on the number under ‘Input 
add time’, it is recommended that the longest available time of 60 seconds is used for the 
most accurate calibration. 

 

Select ‘I’ve finished setting’ at the bottom of the screen to continue. 

 

Place the digital scales on a stable, flat and level surface then power on. 

 

Switch the scales to read in grams by depressing the mode (M) button until ‘g’ is shown. 

 

Remove the glass beaker from the KH Manager and empty out any liquid, then place the 
empty glass beaker on the scales. 

 

Press the Tare button (T) to zero the scales and compensate for the weight of the glass 
beaker then select ‘I’ve tared’ at the bottom of the screen in the app. 

 

Leave the scales on and remove the glass beaker, the display will show the weight of the 
beaker as a minus number. 

 

Place the glass beaker back in the base of the unit under the short tube that is on the far 
right as you look at the unit. 

 

In the app tap ‘Start adding’ to begin running the pump, the screen will show a countdown 
in seconds as the pump runs. 

 

Once the pump stops running tap ‘Add ended’ at the bottom of the screen. 

Summary of Contents for KH Manager

Page 1: ...KH MANAGER INSTRUCTION MANUAL V1 0...

Page 2: ...TO THE CLOUD 7 DIRECT BLUETOOTH CONNECTION TO THE KH MANAGER 7 STARTING THE KH MANAGER 8 IN APP STEP BY STEP GUIDE 8 INSTALING THE FRONT HOUSING 8 MIXING THE KH TEST REAGENT 9 CHECKING THE KH TEST REA...

Page 3: ...R SOLUTION PUMP KH SOLUTION IN 23 CALIBRATING THE KH TEST PUMP KH REAGENT IN 24 CALIBRATING THE KH 25 RESETTING THE CALIBRATION ON THE KH MANAGER 25 MANUALLY RUNNING THE PUMP HEADS 26 RUNNING THE SALT...

Page 4: ...ividually calibrated and easily serviced Data and programming stored on unit in case of power or connection loss Lab grade pH probe DIN port for connection and expansion with other D D Kamoer dosing p...

Page 5: ...e contact your retailer immediately Box Contents Quick start guide KH Manager unit pH Probe Glass sample beaker 5ml Glass measuring cylinder 10ml Measuring cylinder High precision digital scales pH 4...

Page 6: ...ht difference is within 0 5m The unit should be secured into place or installed in a stable location not above open water or in very humid environments The surface must be flat and level to ensure cor...

Page 7: ...stream from where the sample water is being collected this will ensure a fresh sample is being taken for best accuracy when testing If the waste water is being sent to a waste liquid container then ne...

Page 8: ...connection to a dedicated 5G Wi Fi network is not possible Once you have created an account log in and tap on the plus symbol in the middle of the screen if this is the first device being added or on...

Page 9: ...rating specific components is described below section CALIBRATION the unit must be calibrated properly before use INSTALING THE FRONT HOUSING Once the KH Manager has been fully calibrated using the in...

Page 10: ...eagent Supplied with the unit Weigh out exactly 980 grams of pure water RODI Add the complete 20ml bottle of standard concentrate reagent Gently mix the solution Mixing the double strength concentrate...

Page 11: ...and then the blue arrow in the top left corner to exit the Maintain screen If the reagent is out of the acceptable range for use it should be safely discarded and a fresh solution made PERFORMING A T...

Page 12: ...s displaying Automatic tap on the blue downwards arrow and select Manual test from the pop up at the bottom of the screen Next select the menu icon in the top right to enter the settings for the test...

Page 13: ...tion will be added after the test as per the settings To turn off the automatic dosing tap on Tank Volume then tap on the toggle in the top right corner next to Keep KH to move it to the off position...

Page 14: ...xit To turn the function off tap the toggle again and Off will be displayed to indicate it is not active Select the blue arrow in the top left corner to return to the test settings page As default the...

Page 15: ...without losing any settings An immediate test function is also available when the KH Manager is in automatic mode This will allow a one off test without having to switch the machine to manual mode or...

Page 16: ...tap on the Status icon in the bottom left corner A tile will be shown for the KH buffer solution KH Rise Reagent and the KH test reagent that displays the amount of liquid available the starting volum...

Page 17: ...d the setting is shown in the separate KH Manager Expansion instructions There are two options that can be used Auto add and Custom Plans ACTIVATING AND DEACTIVATING AUTOMATIC KH DOSING NOTE The KH ri...

Page 18: ...the pop up that appears on the bottom of the screen The settings on this page let you determine limitations on how much KH buffer solution can be added and how the machine doses the buffer solution T...

Page 19: ...ou need to program the concentration of the solution you are using into the machine To do this tap on KH Rise Reagent Brand in the next page there are two options Kamoer and Other if you are using Kam...

Page 20: ...hen enter a time for the plan to start by selecting Start Time use the scroll wheel to set the time and then save by selecting Done next tap on Add to enter the amount of solution you would like to ad...

Page 21: ...ings page tap on the lower and upper parameters to make adjustments enter the figure then select Done and Save in the top right corner to save and exit Method This is used to determine if you would li...

Page 22: ...ons some pure water RODI for rinsing the probe and some soft paper towel If present remove the protective cap from the pH probe by unscrewing the pot section and then sliding the seal and cap off the...

Page 23: ...rom the KH Manager by pulling it forwards exposing the tubes When prepared tap I m ready in the app Place the glass beaker in the base of the unit under the short tube that is on the far right as you...

Page 24: ...Calibration page will show you details of when the pump head was last calibrated when the next calibration is due this is a user defined setting the current head calibration data and the option to Go...

Page 25: ...enter the number of days then save in the top right corner For accurate testing it is suggested to set this at 45 60 days after this time has elapsed a warning will appear on the screen in the app to...

Page 26: ...sing Please recheck KH If this warning is shown perform a test with the KH Manager before commencing with this calibration Go to the calibration settings in the Kamoer Remote app and select KH from th...

Page 27: ...for that pump to start the pump tap the corresponding box labelled Run the pump will start to add the amount specified and will show a countdown of that amount as it is dosed To stop mid dose tap the...

Page 28: ...select Name from the list then tap on the white space and enter the new name for the unit When finished press Done followed by Save in the top right corner UPDATING THE KH MANAGER FIRMWARE Log into t...

Page 29: ...If you want to reset the calibration settings on the KH Manager go to the main settings tab in the Kamoer Remote app select Maintain the calibration tile and then Reset Calibration tap on Reset and t...

Page 30: ...ship be found within 12 months of the date of purchase D D The Aquarium Solution Ltd undertakes to repair or at our discretion replace the defective part free of charge Our policy is one of continual...

Reviews: