background image

17 

 

needed, so it is crucial that this is entered correctly

. To set this tap on ‘Tank Volume’ enter the new 

value (in litres) on the number pad that appears and confirm by selecting ‘Ok’, if you want to exit 
without changing the value choose ‘Cancel’. 

Tracking is also shown on this page for water that has been used for the testing process, under the 
heading ‘Seawater Used’, this is purely a tracking element and can be viewed for reference. If at any 
point you want to clear this tracking element and set it back to zero tap on ‘Seawater Used’ followed 
by ‘Recalculate’ in the pop-up that appears at the bottom of the screen, if you want to exit without 
clearing select ‘Cancel’. 

 

SETTING AUTO ADD MODE 

After each KH test in this mode the machine will calculate how much KH buffer solution will need to 
be added to raise the level in your aquarium to the set point. For example if your KH set point (KH 
Balance Value) is 7.5dKH and the test has a result of 7.3dKH then enough KH buffer will be added to 
raise your aquarium by 0.2dKH. If the result is on or higher than your set point, then no buffer will be 
added.  

To activate and set this mode go to the ‘KH Status’ page as described above under the heading 
‘ACTIVATING AND DEACTIVATING AUTOMATIC KH DOSING’, then select ‘Add KH Settings’. The next 
screen will show different device options for adding the KH buffer, choose ‘Use “KH Carer” to add’. 
This will take you to the settings page for dosing using the KH Manager, at the top ‘Auto Add’ or 
‘Custom Plans’ will be displayed. If ‘Custom Plans’ is shown, tap on the blue downwards facing arrow 
at the end of the heading and select ‘Auto Add’ from the pop-up that appears on the bottom of the 
screen. 

The settings on this page let you determine limitations on how much KH buffer solution can be 
added and how the machine doses the buffer solution.  

Total add cap:

 This is the maximum total amount of KH buffer that the machine will be allowed to 

dose after a test. If the KH value in the aquarium has dropped significantly for some reason this will 
prevent too much buffer solution from being added in one go and allow the levels to be returned to 
normal gradually.  

Generally, we would suggest that the KH elevation per day is limited to 0.5dKH, to accurately 
calculate the figure to set on an established aquarium you will also need to take into account the 
amount of KH buffer your aquarium will normally use per day. So for example if the KH Manager is 
performing 4 tests per day, it takes 10ml of KH buffer to raise your aquarium by the 0.5dKH limit and 
you normally add 10ml daily to the aquarium to maintain the KH level the total add cap should be 
set to 5ml (20ml ÷ 4 tests = 5ml). To adjust the amount, tap on ‘Total Add Cap’, then on the white 
box at the top of the page, enter the amount in ml followed by ‘Done’, finally tap save in the top 
right corner to confirm and return to the settings page. 

Single add cap: 

The KH Manager can break up the

 

calculated automatic dose into smaller amounts 

to reduce the risk of precipitation and give a more gradual correction to the KH level in the 
aquarium. For example, the machine has calculated that it needs to add 4ml of KH buffer solution 
after a test to correct the KH level, if you have the single add cap set to 1ml then it will add four 
doses of 1ml (1 + 1 + 1 + 1 = 4ml). The gap between the doses is determined by the KH interval 
described in the next paragraph. If the dose amount needed does not work out exactly as the 

Summary of Contents for KH Manager

Page 1: ...KH MANAGER INSTRUCTION MANUAL V1 0...

Page 2: ...TO THE CLOUD 7 DIRECT BLUETOOTH CONNECTION TO THE KH MANAGER 7 STARTING THE KH MANAGER 8 IN APP STEP BY STEP GUIDE 8 INSTALING THE FRONT HOUSING 8 MIXING THE KH TEST REAGENT 9 CHECKING THE KH TEST REA...

Page 3: ...R SOLUTION PUMP KH SOLUTION IN 23 CALIBRATING THE KH TEST PUMP KH REAGENT IN 24 CALIBRATING THE KH 25 RESETTING THE CALIBRATION ON THE KH MANAGER 25 MANUALLY RUNNING THE PUMP HEADS 26 RUNNING THE SALT...

Page 4: ...ividually calibrated and easily serviced Data and programming stored on unit in case of power or connection loss Lab grade pH probe DIN port for connection and expansion with other D D Kamoer dosing p...

Page 5: ...e contact your retailer immediately Box Contents Quick start guide KH Manager unit pH Probe Glass sample beaker 5ml Glass measuring cylinder 10ml Measuring cylinder High precision digital scales pH 4...

Page 6: ...ht difference is within 0 5m The unit should be secured into place or installed in a stable location not above open water or in very humid environments The surface must be flat and level to ensure cor...

Page 7: ...stream from where the sample water is being collected this will ensure a fresh sample is being taken for best accuracy when testing If the waste water is being sent to a waste liquid container then ne...

Page 8: ...connection to a dedicated 5G Wi Fi network is not possible Once you have created an account log in and tap on the plus symbol in the middle of the screen if this is the first device being added or on...

Page 9: ...rating specific components is described below section CALIBRATION the unit must be calibrated properly before use INSTALING THE FRONT HOUSING Once the KH Manager has been fully calibrated using the in...

Page 10: ...eagent Supplied with the unit Weigh out exactly 980 grams of pure water RODI Add the complete 20ml bottle of standard concentrate reagent Gently mix the solution Mixing the double strength concentrate...

Page 11: ...and then the blue arrow in the top left corner to exit the Maintain screen If the reagent is out of the acceptable range for use it should be safely discarded and a fresh solution made PERFORMING A T...

Page 12: ...s displaying Automatic tap on the blue downwards arrow and select Manual test from the pop up at the bottom of the screen Next select the menu icon in the top right to enter the settings for the test...

Page 13: ...tion will be added after the test as per the settings To turn off the automatic dosing tap on Tank Volume then tap on the toggle in the top right corner next to Keep KH to move it to the off position...

Page 14: ...xit To turn the function off tap the toggle again and Off will be displayed to indicate it is not active Select the blue arrow in the top left corner to return to the test settings page As default the...

Page 15: ...without losing any settings An immediate test function is also available when the KH Manager is in automatic mode This will allow a one off test without having to switch the machine to manual mode or...

Page 16: ...tap on the Status icon in the bottom left corner A tile will be shown for the KH buffer solution KH Rise Reagent and the KH test reagent that displays the amount of liquid available the starting volum...

Page 17: ...d the setting is shown in the separate KH Manager Expansion instructions There are two options that can be used Auto add and Custom Plans ACTIVATING AND DEACTIVATING AUTOMATIC KH DOSING NOTE The KH ri...

Page 18: ...the pop up that appears on the bottom of the screen The settings on this page let you determine limitations on how much KH buffer solution can be added and how the machine doses the buffer solution T...

Page 19: ...ou need to program the concentration of the solution you are using into the machine To do this tap on KH Rise Reagent Brand in the next page there are two options Kamoer and Other if you are using Kam...

Page 20: ...hen enter a time for the plan to start by selecting Start Time use the scroll wheel to set the time and then save by selecting Done next tap on Add to enter the amount of solution you would like to ad...

Page 21: ...ings page tap on the lower and upper parameters to make adjustments enter the figure then select Done and Save in the top right corner to save and exit Method This is used to determine if you would li...

Page 22: ...ons some pure water RODI for rinsing the probe and some soft paper towel If present remove the protective cap from the pH probe by unscrewing the pot section and then sliding the seal and cap off the...

Page 23: ...rom the KH Manager by pulling it forwards exposing the tubes When prepared tap I m ready in the app Place the glass beaker in the base of the unit under the short tube that is on the far right as you...

Page 24: ...Calibration page will show you details of when the pump head was last calibrated when the next calibration is due this is a user defined setting the current head calibration data and the option to Go...

Page 25: ...enter the number of days then save in the top right corner For accurate testing it is suggested to set this at 45 60 days after this time has elapsed a warning will appear on the screen in the app to...

Page 26: ...sing Please recheck KH If this warning is shown perform a test with the KH Manager before commencing with this calibration Go to the calibration settings in the Kamoer Remote app and select KH from th...

Page 27: ...for that pump to start the pump tap the corresponding box labelled Run the pump will start to add the amount specified and will show a countdown of that amount as it is dosed To stop mid dose tap the...

Page 28: ...select Name from the list then tap on the white space and enter the new name for the unit When finished press Done followed by Save in the top right corner UPDATING THE KH MANAGER FIRMWARE Log into t...

Page 29: ...If you want to reset the calibration settings on the KH Manager go to the main settings tab in the Kamoer Remote app select Maintain the calibration tile and then Reset Calibration tap on Reset and t...

Page 30: ...ship be found within 12 months of the date of purchase D D The Aquarium Solution Ltd undertakes to repair or at our discretion replace the defective part free of charge Our policy is one of continual...

Reviews: