background image

14 

 

It is also possible for the machine to only carry out the test programs on certain days of the week or 
every number of ‘X’ days. The current setting is displayed at the bottom of the automatic test page 
screen under ‘Cycle Date’. To change the days the KH Manager will perform tests select ‘Cycle Date’ 
at the bottom of the screen. A pop-up will appear with the options ‘By Days’ and ‘Weekly’.  
 

 

Select ‘By days’ to set the tests every ‘x’ days, i.e. if set to 2 then the machine will perform 
the tests every 2 days. Choose the frequency using the scroll wheel at the bottom of the 
page and then tap ‘Save’ in the top right corner. 

 

Select ‘Weekly’ to choose the days of the week the KH Manager will perform the tests, i.e. 
you could choose only to test on a Monday, Wednesday, and Friday. Tap on the circle to the 
right of each day to select or de-select it, any days showing a corresponding highlighted 
circle with a tick in it are the days it will test. Once done tap ‘Save’ in the top right corner. 

 
On the automatic test page, above the plan times, there is a box that will be displaying either 
‘Testing Plans : OFF’ or ‘Testing Plans : ON’. When ‘OFF’ all the programming will remain stored but 
the machine will not run, if ‘ON’ the programs will be active and the KH Manager will test as set. To 
switch between on and off tap on the toggle switch to the right, when the toggle is highlighted it will 
switch to on and when grey it will be off. The off function can be used during machine or aquarium 
maintenance to deactivate it without losing any settings. 
 
An immediate test function is also available when the KH Manager is in automatic mode. This will 
allow a one-off test without having to switch the machine to manual mode or alter any settings, 
when the test is complete it will return to automatic testing as set. To perform an immediate test, 
select ‘Test Now’, located under the list of plan times in the automatic test page, then confirm by 
tapping on ‘Test’ in the pop-up that appears at the bottom of the screen. To cancel and exit tap on 
the screen anywhere above the pop-up. 
 

Please note: An automatic test will be skipped if an immediate test overlaps its start time or you 
are in the ‘Maintain’ section within the main settings tab, the next test program will continue as 
normal. 
 

DATA LOGGING AND MONITORING 

If your KH Manager is registered to an account on the Kamoer Remote app and connected to the 
cloud via Wi-Fi it will record data after each test. Data is collected on the KH test results, pH of the 
tank water sample, time the test was completed, the test mode used and depending on the dosing 
mode how much KH Buffer solution is added (If automatic dosing is activated using the KH 
Manager).  

To view the collected data, log into the Kamoer Remote app and tap on the device icon for your KH 
Manager, then select the ‘Data’ icon from the bottom of the screen. The page will display a graph 
plotting the results from each test at the top and a data list below.  

Two separate data pages are available one for KH and one for pH. To select the data you would like 
to view tap on either the KH or pH heading at the top of the data page, a bar highlighting the data 
you are viewing will be shown underneath the KH or pH heading. 

Summary of Contents for KH Manager

Page 1: ...KH MANAGER INSTRUCTION MANUAL V1 0...

Page 2: ...TO THE CLOUD 7 DIRECT BLUETOOTH CONNECTION TO THE KH MANAGER 7 STARTING THE KH MANAGER 8 IN APP STEP BY STEP GUIDE 8 INSTALING THE FRONT HOUSING 8 MIXING THE KH TEST REAGENT 9 CHECKING THE KH TEST REA...

Page 3: ...R SOLUTION PUMP KH SOLUTION IN 23 CALIBRATING THE KH TEST PUMP KH REAGENT IN 24 CALIBRATING THE KH 25 RESETTING THE CALIBRATION ON THE KH MANAGER 25 MANUALLY RUNNING THE PUMP HEADS 26 RUNNING THE SALT...

Page 4: ...ividually calibrated and easily serviced Data and programming stored on unit in case of power or connection loss Lab grade pH probe DIN port for connection and expansion with other D D Kamoer dosing p...

Page 5: ...e contact your retailer immediately Box Contents Quick start guide KH Manager unit pH Probe Glass sample beaker 5ml Glass measuring cylinder 10ml Measuring cylinder High precision digital scales pH 4...

Page 6: ...ht difference is within 0 5m The unit should be secured into place or installed in a stable location not above open water or in very humid environments The surface must be flat and level to ensure cor...

Page 7: ...stream from where the sample water is being collected this will ensure a fresh sample is being taken for best accuracy when testing If the waste water is being sent to a waste liquid container then ne...

Page 8: ...connection to a dedicated 5G Wi Fi network is not possible Once you have created an account log in and tap on the plus symbol in the middle of the screen if this is the first device being added or on...

Page 9: ...rating specific components is described below section CALIBRATION the unit must be calibrated properly before use INSTALING THE FRONT HOUSING Once the KH Manager has been fully calibrated using the in...

Page 10: ...eagent Supplied with the unit Weigh out exactly 980 grams of pure water RODI Add the complete 20ml bottle of standard concentrate reagent Gently mix the solution Mixing the double strength concentrate...

Page 11: ...and then the blue arrow in the top left corner to exit the Maintain screen If the reagent is out of the acceptable range for use it should be safely discarded and a fresh solution made PERFORMING A T...

Page 12: ...s displaying Automatic tap on the blue downwards arrow and select Manual test from the pop up at the bottom of the screen Next select the menu icon in the top right to enter the settings for the test...

Page 13: ...tion will be added after the test as per the settings To turn off the automatic dosing tap on Tank Volume then tap on the toggle in the top right corner next to Keep KH to move it to the off position...

Page 14: ...xit To turn the function off tap the toggle again and Off will be displayed to indicate it is not active Select the blue arrow in the top left corner to return to the test settings page As default the...

Page 15: ...without losing any settings An immediate test function is also available when the KH Manager is in automatic mode This will allow a one off test without having to switch the machine to manual mode or...

Page 16: ...tap on the Status icon in the bottom left corner A tile will be shown for the KH buffer solution KH Rise Reagent and the KH test reagent that displays the amount of liquid available the starting volum...

Page 17: ...d the setting is shown in the separate KH Manager Expansion instructions There are two options that can be used Auto add and Custom Plans ACTIVATING AND DEACTIVATING AUTOMATIC KH DOSING NOTE The KH ri...

Page 18: ...the pop up that appears on the bottom of the screen The settings on this page let you determine limitations on how much KH buffer solution can be added and how the machine doses the buffer solution T...

Page 19: ...ou need to program the concentration of the solution you are using into the machine To do this tap on KH Rise Reagent Brand in the next page there are two options Kamoer and Other if you are using Kam...

Page 20: ...hen enter a time for the plan to start by selecting Start Time use the scroll wheel to set the time and then save by selecting Done next tap on Add to enter the amount of solution you would like to ad...

Page 21: ...ings page tap on the lower and upper parameters to make adjustments enter the figure then select Done and Save in the top right corner to save and exit Method This is used to determine if you would li...

Page 22: ...ons some pure water RODI for rinsing the probe and some soft paper towel If present remove the protective cap from the pH probe by unscrewing the pot section and then sliding the seal and cap off the...

Page 23: ...rom the KH Manager by pulling it forwards exposing the tubes When prepared tap I m ready in the app Place the glass beaker in the base of the unit under the short tube that is on the far right as you...

Page 24: ...Calibration page will show you details of when the pump head was last calibrated when the next calibration is due this is a user defined setting the current head calibration data and the option to Go...

Page 25: ...enter the number of days then save in the top right corner For accurate testing it is suggested to set this at 45 60 days after this time has elapsed a warning will appear on the screen in the app to...

Page 26: ...sing Please recheck KH If this warning is shown perform a test with the KH Manager before commencing with this calibration Go to the calibration settings in the Kamoer Remote app and select KH from th...

Page 27: ...for that pump to start the pump tap the corresponding box labelled Run the pump will start to add the amount specified and will show a countdown of that amount as it is dosed To stop mid dose tap the...

Page 28: ...select Name from the list then tap on the white space and enter the new name for the unit When finished press Done followed by Save in the top right corner UPDATING THE KH MANAGER FIRMWARE Log into t...

Page 29: ...If you want to reset the calibration settings on the KH Manager go to the main settings tab in the Kamoer Remote app select Maintain the calibration tile and then Reset Calibration tap on Reset and t...

Page 30: ...ship be found within 12 months of the date of purchase D D The Aquarium Solution Ltd undertakes to repair or at our discretion replace the defective part free of charge Our policy is one of continual...

Reviews: