ANNEX II: UPGRADING FIRMWARE AND LOADING PRESETS (cont'd)
Start The Update!
Just before you press the “Load” button – please close any other applications running on the
PC – interruptions in the process could render your amplifier inoperable. Do not disconnect the
cable, turn off the power or attempt to close the loader app once loading has begun!
Now that you have been warned (!) – press the “Load” button. Confirm the load, and the amplifier
will switch to Standby and begin to load the new firmware. A message detailing progress will accompany
the bar on the loader app:
Upon completion, the app will close and your amplifier will restart – that's it!
What If It All Goes Wrong…
Don't panic! Your amplifier can be recovered, but please be aware that…
THE AMPLIFIER OPERATES AT VOLTAGES IN EXCESS OF THE INCOMING MAINS VOLTAGE,
AND THESE VOLTAGES REMAIN PRESENT FOR AN EXTENDED PERIOD AFTER MAINS POWER IS
REMOVED!
READ THESE INSTRUCTIONS CAREFULLY AND HEED ALL WARNINGS!
Disconnect your amplifier completely from the incoming mains supply.
Ensure that the HT present LEDs visible here through the front left vent with the foam removed…
Remote Load 374B
Block 0 of 4096
…are fully extinguished before continuing. This will take at least five minutes so please be patient for
your own safety.