![Cynosure Acclaim 7000 Technical Manual Download Page 53](http://html1.mh-extra.com/html/cynosure/acclaim-7000/acclaim-7000_technical-manual_2704653053.webp)
Apogee Elite/Apogee 5500/Acclaim 7000
850-1257-000, Rev. 2
Page 49 of 70
Technical Guide
Section 9
SERVICE PROCEDURES–ELECTRICAL
For most electrical procedures involving the replacement, recalibration and/or test of a
component, refer to the
Apogee Elite/Apogee 5500/Acclaim 7000 Final Test and Calibration
Procedure
for complete instructions.
Replacing the Front Panel Display/Membrane Switch PCB
1.
Turn off the laser power and disconnect the AC from the wall.
2.
Remove four screws securing the front panel display to the top control/display plate.
3.
Remove the two wires form the emergency stop switch.
4.
Disconnect P13 and fiber #97 for the cal port assembly.
5.
Remove all connectors from the membrane switch PCB and the display PCB.
6.
Using a nut driver, remove the four 4-40 nuts that hold the display PCB, or the eight 4-40
nuts that hold the membrane switch PCB to the front panel plate.
7.
Replace the PCB by reversing the above steps.
Replacing and Calibrating the Control Logic PCB
1.
Turn off the laser power and disconnect the AC from the wall.
2.
Follow the section above to remove the front panel display from the top control plate.
3.
Remove all connectors from the PCB. Connectors should be marked, but pin number and
harness length should only allow one possibility for connection.
4.
Remove the six 4-40 screws that attach the PCB to the sheet metal.
5.
Replace the PCB by reversing the above steps.
6.
Perform the following sections of the
Final Test and Calibration Procedure
:
♦
HVPS Calibration
♦
Temperature Settings
♦
Calibrating the Energy Monitor
♦
Calibrating the Cal Port
Programming the Microcontroller
For updating laser software, follow the instructions found in the
Apogee 5500/Acclaim7000
Software Update Procedure Using a Laptop Computer
.
Summary of Contents for Acclaim 7000
Page 2: ......
Page 3: ...Apogee Elite Apogee 5500 Acclaim 7000 Long Pulse Infrared Lasers TECHNICAL GUIDE...
Page 10: ...Page 6 of 70 850 1257 000 Rev 2 Cynosure Inc This page has been intentionally left blank...
Page 18: ...Page 14 of 70 850 1257 000 Rev 2 Cynosure Inc Figure 2B Block Diagram Apogee 5500 Acclaim 7000...
Page 24: ...Page 20 of 70 850 1257 000 Rev 2 Cynosure Inc Figure 6 IGBT Plate Assembly...
Page 28: ...Page 24 of 70 850 1257 000 Rev 2 Cynosure Inc Figure 8A Resonator Rail Apogee Elite...
Page 36: ...Page 32 of 70 850 1257 000 Rev 2 Cynosure Inc This page has been intentionally left blank...
Page 38: ...Page 34 of 70 850 1257 000 Rev 2 Cynosure Inc This page has been intentionally left blank...
Page 44: ...Page 40 of 70 850 1257 000 Rev 2 Cynosure Inc This page has been intentionally left blank...
Page 52: ...Page 48 of 70 850 1257 000 Rev 2 Cynosure Inc This page has been intentionally left blank...
Page 56: ...Page 52 of 70 850 1257 000 Rev 2 Cynosure Inc This page has been intentionally left blank...
Page 60: ...Page 56 of 70 850 1257 000 Rev 2 Cynosure Inc This page has been intentionally left blank...
Page 66: ...Page 62 of 70 850 1257 000 Rev 2 Cynosure Inc This page has been intentionally left blank...
Page 72: ...Page 68 of 70 850 1257 000 Rev 2 Cynosure Inc This page has been intentionally left blank...
Page 74: ...Page 70 of 70 850 1257 000 Rev 2 Cynosure Inc This page has been intentionally left blank...