5. Machinery Specification
Type
Parameter
Differential input cable
15cm
BNC output cable (CK-310)
1m
Alligator clips (CK-261)
85*40*17mm
Piston clips (CK-281)
152*50*13mm
Test hook (CK-284)
121*23*23mm
Probe dimension
Front
116*22*15mm
Back
124*49*27mm
Probe weight
172g
6. Environmental Characteristics
Type
Parameter
Pollution level
2
Operating temperature
0
℃~
50
℃
Storage temperature
-30
℃~
70
℃
Operating humidity
≤85%RH
Storage humidity
≤90%RH
Operating altitude
3000m
Storage altitude
12000m
7. Operating Steps
The range of the voltage under test should be estimated before the test. Voltage surpassing the
measuring range could possibly damage the probe and the product.
Connect the input and output cable to the probe and connect the probe with oscilloscope or other test
devices.
The green power indicator will be lighted when the power supply adaptor connects with the voltage
probe. When the voltage under test surpass the range, the indicator will be lighted along with the
buzzer.
Set the attenuation ratio of the oscilloscope or other devices to 10:1, input impedance to 50Ω (If the
probe output port is connected with 50Ω thorough type load, the input impedance should be 1MΩ)
Adjust the sensitivity of the oscilloscope according to the voltage under test
Connect the clips according to needs and start the test. The main body of the probe should be put
away from the high voltage pulse circuit to minimize the interference.
Turn off the power of the circuit under test after the test is over, then turn off the power of the probe
and disconnect the two input ports from the point under test, and unplug the BNC connector from
oscilloscope.