8.Error messages/troubleshooting
If an error occurs during a measurement, the measurement is discontinued
and a corresponding error code is displayed (example: Error no. 2).
Error No.
Possible cause(s)/Solutions
ERR 1
The tube may have loosened, or no pulse was detected.* Ensure
cuff connections are tight with proper cuff placement. See section 5.3.
ERR 2 Unnatural pressure impulses influenced the measurement result.
Reason: The arm was moved during the measurement (artefact).
Repeat measurement, keeping still and quiet.
ERR 3 Inflation of the cuff takes too long. The cuff is not correctly seated
or the hose connection is not tight. Re-position cuff and repeat the
measurement.
ERR 5 The measured readings indicated an unacceptable difference
between systolic and diastolic pressures. Take another reading
following directions carefully. Contact your doctor if you continue to