7.
Move the handpiece to the next area to be treated with an overlap of 20-50%, as shown
below.
Completing a Session
At the end of the treatment session:
1.
Place the system in STANDBY mode, and store the RF handpiece in the handpiece
cradle.
2.
Lift the latch on the neutral electrode cable, and remove the return pad tab from the clamp.
3.
Slowly remove the return pad from the patient’s skin.
4.
Dispose of the return pad.
5.
Disinfect the RF handpiece, as described in the Maintenance chapter of this manual.
6.
To begin a new treatment, touch the “New Treatment” button on the Select screen. To
review the treatment summary, touch the “Resume Treatment” button on the Select screen
to return to the RF Handpiece Control screen, and then touch the
tab to go to the
Information & Adjustment screen.
CAUTION
The neutral electrode cable must be disconnected from the return
pad BEFORE removing the pad from the patient.
Correct RF Handpiece Placement
Incorrect RF Handpiece Placement