15
TRUSCULPT OPERATOR MANUAL
D1395, REV. K, 02/18
Remaining Uses Indicator
The Remaining Uses indicator displays the number of RF treatments remaining before handpiece
replacement is required. The truSculpt 16 and truSculpt 40 are limited to 51 uses. When five uses
remain, the indicator changes from green to yellow. When zero uses remain, the “New Treatment” button
is disabled, and you must replace the handpiece. You can request handpiece replacement through the
Cutera website (www.cutera.com/onlinestore).
Location of Remaining Uses Indicator on Select Screen
NOTE
Each RF treatment consists of 64 minutes “RF on” time, or 90
minutes total treatment time (to allow for breaks). The total
treatment time starts when you press the RF handpiece on/off
button and continues until either 90 minutes has elapsed or the
“New Treatment” button is pressed. Turning off the system does not
reset the total treatment time.
Remaining Uses
Indicator