KAD/ADC/129/S2
6
Setting up the KAD/ADC/129/S2
All module setup can be defined in XML using XidML® schemas (see
).
Instrument settings
Parameter definitions
Configurable parameters
AnalogIn(3:0)
N
OTE
:
It is recommended that names are less than 20 characters, have no white space or contain any of the following five char
-
acters "/><\.
SETUP DATA
CHOICE
DEFAULT
NOTES
Manufacturer
-
-
-
Name
ACRA CONTROL
ACRA CONTROL
Name of manufacturer.
PartReference
KAD/ADC/129/S2
KAD/ADC/129/S2
The instrument part reference.
SerialNumber
AB1234
AB1234
Unique name for each module.
Channels
-
-
-
AnalogIn(3:0)
Analog Input
-
-
-
Settings
-
-
-
Filter Cutoff
0.25
0.5
1
2
4
8
16
0.25
Required cutoff point for the filter is the chosen value
multiplied by the user sampling frequency. 0.25 is
recommended as any higher may lead to aliasing. 1
is the sampling rate.
Excitation Amplitude
0 to 5.1
0.2
Required excitation (in V) for the top of the bridge.
Excitation is bipolar so entering 5V means 10V
across the bridge.
Balance.Type
CurrentShunt
CurrentShunt
Specifies the balance type to be carried out on the
bridge.
Balance.Applied
-71e-6 to 71e-6
0
Shunt current (in A) applied to the bridge.
Balance.BalanceThisTime
True
False
False
Specifies if balancing should be carried out this time
by software.
Balance.Tolerance
0.01 to 99.99
0.1
Specifies acceptable tolerance of achieved value vs.
target value, expressed as percentage of defined
input range.
Balance.Target
-10 to 10
0
Specifies a value. that the channel should be
balanced to.
ShuntCurrent.Applied
-71e-6 to 71e-6
0
Shunt mode current (in A) added to the bridge
NAME/DESCRIPTION
BASE UNIT
DATA FORMAT
BITS
REGISTER DEFINITION
AnalogIn(3:0) Parameters
AnalogIn
Analog signal data
Volt
OffsetBinary
16
R[15:0]
SETUP DATA
CHOICE
DEFAULT
NOTES
Range Maximum
-10 to 10
10
Range maximum for analog channel
Range Minimum
-10 to 10
-10
Range minimum for analog channel