M869-GB-2_Letter
9 / 10
mini NEB US
5.3. Inspection and Cleaning of the Discharge Siphon
Pic
Pic
Pic
Pic. . . . 5.3
5.3
5.3
5.3....1
11
1 ---- Siphon
Siphon
Siphon
Siphon
It may be necessary to clean periodically the water
discharge siphon R (see: Pic. 5.3.1) as the
accumulation of dirt inside may compromise its correct
and efficient operation. When a cleaning is necessary,
see the following steps:
•
Extract the hose T from the small pipe R;
•
Take off the screws V;
•
Take off the small pipe R;
•
Carefully clean both the small pipe R and the pipe
S, that is located inside the basin;
Once cleaned, reassemble carefully all the pieces.
5.4. Inspection and cleaning of the water feed solenoid valve
The water feed solenoid valve is fitted with an inlet
filter that must be periodically checked and cleaned.
To access the filter, unscrew the plastic threaded bush
A of the water supply hose: the filter is located inside
the threaded bush E of the solenoid valve.
If cleaning becomes too frequent, it is advisable to
install a cartridge filter on the machine water supply
line(see: paragraph 3.4 - Hydraulic connections)
Pic
Pic
Pic
Pic. . . . 5.4
5.4
5.4
5.4....1
11
1 –––– Solenoid valve
Solenoid valve
Solenoid valve
Solenoid valve
6. Storage
6.1. General
notes
•
Store the machine in a room with a temperature between 14 °F and 140 °F.
•
When the machine is in its package, keep the package vertical.
•
Do not stack other packages on top.
6.2. Verifications to be made before and after a long working period
6.2.1. Before
•
Disconnect all electrical connections and close water feed supply tap.
•
Cover the machine to protect it from dust
6.2.2. After
•
Check the air intake filter, clean it if necessary.
•
Check that the floating switch works correctly, moving it, and check that the fan/disk assembly rotates
freely.
•
Open the water supply tap.
•
Check that all electrical connections are correctly executed, as by instructions.