background image

12

TARGET BIN SCALE RF rev.09

TARGET BIN SCALE RF

5.6.  Bin Calibration

It is possible to calibrate bin scales using one 
or 2 reference weights:

1.  The  1-point  calibration  mode  allows 
calibrating  a  bin  scale  with  only  one  refer-
ence weight: the weight of the feed that is 

currently present in the scale. This type of 
calibration can be performed at any time. 

2.  The 2-point calibration mode allows using 
two reference weights: the empty bin weight 
and the loaded bin weight; the bin has to be 
empty prior to the calibration.

5.6.1.  1-Point Calibration

The 1-point calibration can be performed at any 
time, as long as you know the amount of feed 
present in the bin. If you don't know what is 
the exact amount of feed, you can estimate it: 
in that case, disable the "Low Level" option in 
section 5.5 and be sure to re-calibrate the scale 
as soon as the bin gets empty.

Repeat the 1-point calibration each time 
the scale gets empty and its weight read-
ing is other than 0 kg (or 0 lbs).

1.  From the main menu, select:

 

»   7. Installation*

 

»   2. Calibration.

 

»    x. Select the desired bin

 

»    1. Bin #x One Point

*This  menu  is  only  accessible  from  the  installer 

mode (see sec. 5.2).

 

7.2.1 Calibr Bin 1
Nb Load Cells : 

4

Sen. LC1 : 

3.0000

Sen. LC2 : 

3.0000

Sen. LC3 : 

3.0000

Sen. LC4 : 

3.0000

Max Cap. : 10000 lbs
        [or 4536 kg]
Tare     : 

0 lbs

Offset: 

68764115

Stability tol: 2000
Push + for next step
Calibrate ?  No

2.  Set the following parameters:

Number of Load Cells: —

 Select the number 

of load cells that are supporting the bin (from 
1 to 8 load cells).

Sensitivity of Load Cells —

 Enter the sensi-

tivity of each load cell. The sensitivity value 
is specific to each load cell and is indicated 
on the load cell's calibration certificate. It is 
also written on each one of them.

Maximum  Capacity  —

  Enter  the  maximum 

weight capacity of load cells. This value is 
indicated  on  the  calibration  certificate  en-
closed with each load cell and it is written 
on each one of them. The maximum capacity 
is common to all load cells. 

The maximum capacity is often given in 
pounds. If your controller uses the metric 
system, be sure to convert this value 
properly. For instance, 10,000 pounds 
corresponds to 4,536 kilograms. 

Tare—

 Specify the amount of feed that is cur-

rently present in the bin scale (do not include 
the weight of the bin).

Offset —

 The offset is a parameter that al-

lows you to recover your calibration settings 
after installing a new configuration in your 
controller. This value is automatically defined 
by the controller after having calibrated the 
scale. 

29

TARGET BIN SCALE RF rev.09

TARGET BIN SCALE RF

Voltage checkup

LCT-3

RED

GREEN

WHITE

SHD

BLK

WHT

GRN

RED

BLACK

B

LK

W

H

T

G

R

N

R

E

D

B

LK

W

H

T

G

R

N

R

E

D

L1

V

REF

J9

V

IN

V

OUT

JX

LCJB-8

Load Bar

V

div2

V

div1

V

REF

 reading on the LCJB-8 module

Param-

eter

Specification

Load 

bar 

1

Load 

bar 

2

Load 

bar 

3

Load 

bar 

4

Load 

bar 

5

Load 

bar 

6

Load 

bar 

7

Load 

bar 

8

Input 

voltage  

(V

in

)

V

IN

    

V

REF

V

REF

 should be between 

4.3 and 5V, depend-

ing on the gauge and 

length (L1) of the wire

Output 

voltage 

(V

out

)

15mV max 

V

out

 should be between 

0.0 and 15.0mV de-

pending on the weight 

of the bin

Divided 

Voltage 

(V

div1

)

V

div1

 should be 

close to V

REF

/2

Divided 

Voltage 

(V

div2

)

V

div2

 should be 

close to V

REF

/2

Summary of Contents for Target

Page 1: ...M 890 00448 rev 09 REV 08 Target Bin Scale RF User s manual...

Page 2: ...be void if this product is used in a manner not specified by the manu facturer Every effort has been made to ensure that this manual is complete accurate and up to date The information contained in i...

Page 3: ...T BIN SCALE RF rev 09 TARGET BIN SCALE RF TABLE OF CONTENTS 1 INTRODUCTION 4 1 1 Precautions 4 1 2 Symbols of the Manual 4 2 SYSTEM OVERVIEW 5 2 1 Description of the Controller 5 2 2 Typical Applicati...

Page 4: ...al earth terminals which are generally asso ciated with test and measurement circuits These terminals are not for safetyearthingpurposesbutprovide an earth reference point Fuse For Customer Use Enter...

Page 5: ...ell Ground strap bolted to the bin leg The bin leg should never touch the load cell it must rest on the washers 5 TARGET BIN SCALE RF rev 09 TARGET BIN SCALE RF 2 SYSTEM OVERVIEW 2 1 Description of th...

Page 6: ...1 Auger 3 Auger 4 Auger 2 Receiving Hopper 35 TARGET BIN SCALE RF rev 09 TARGET BIN SCALE RF Time Clock Settings Refer to section 6 2 3 to program time clocks and refer to section 5 4 to assign them t...

Page 7: ...o be completely opened right up against the wall 3 2 Connections 3 2 1 Controller s Main Wiring Refer to the wiring diagram enclosed with this user s manual to connect the controller Use the electrica...

Page 8: ...a menu has more than two items the display scrolls to show additional items Use the right arrow key to select a menu option Use the left arrow key to return to the previous menu Note that if passwords...

Page 9: ...shes on the display Use the adjustment buttons to set the day 6 Press the right arrow key The hour flashes on the display Use the adjustment buttons to set the hour 7 Proceed in similar fashion to set...

Page 10: ...meter is accessible if the clock schedule is set to Week above Clock Type A time clock can either stop after a specific run time start run or at a pre cise time of the day start stop Choose the desire...

Page 11: ...rs Re fer to section 6 2 3 to program the timer Worksheets are provided at the end of this manual to write down your time clock relay settings 1 From the main menu select 7 Installation 5 Feed Line Se...

Page 12: ...cells Sensitivity of Load Cells Enter the sensi tivity of each load cell The sensitivity value is specific to each load cell and is indicated on the load cell s calibration certificate It is also writ...

Page 13: ...in when the weight reading is stable enough Otherwise it will display an error message The stability tolerance is the variation in the weight reading the controller can tolerate to calibrate the bin s...

Page 14: ...se the control ler displays an error message The stability tolerance is the stability value the controller can tolerate to calibrate the scale By default it is set to 2000 The weight reading may be un...

Page 15: ...ate Maximum State Select Yes to calibrate the loaded bin The message Please Wait is displayed Wait until the end of the calibration this process may take a few minutes If calibration is successful the...

Page 16: ...topping augers when the system is load ed Bin augers automatically stop when the feeding system is full the proximity sensor located at the top of the receiving hopper detects the presence of feed Sto...

Page 17: ...ound today Set these values separately for males and for females 1 From the main menu select 3 House 1 Flock info x Select the desired room x Select the desired type of birds 2 Set the following param...

Page 18: ...justed from one day out of 2 to 8 days Set the frequency to the desired value Days Left This is the number of days that are left before the next skipped day If required you can modify the number of da...

Page 19: ...100 75g Today Tot 37g Last 100 0g Last Tot 0g Remaining Birds 50 2 Set the following parameters The parameters below will be displayed if the feed schedule is set to Skip in the Installation menu sec...

Page 20: ...Select On to activate the auger select Off to stop it or select Auto to go back to the automatic control mode The auger can only be activated manually if its bin is not cur rently sending feed to any...

Page 21: ...Select On to activate the auger select Off to stop it or select Auto to go back to the automatic control mode The auger can only be activated manually if its bin is not cur rently sending feed to any...

Page 22: ...100 75g Today Tot 37g Last 100 0g Last Tot 0g Remaining Birds 50 2 Set the following parameters The parameters below will be displayed if the feed schedule is set to Skip in the Installation menu sec...

Page 23: ...justed from one day out of 2 to 8 days Set the frequency to the desired value Days Left This is the number of days that are left before the next skipped day If required you can modify the number of da...

Page 24: ...ound today Set these values separately for males and for females 1 From the main menu select 3 House 1 Flock info x Select the desired room x Select the desired type of birds 2 Set the following param...

Page 25: ...topping augers when the system is load ed Bin augers automatically stop when the feeding system is full the proximity sensor located at the top of the receiving hopper detects the presence of feed Sto...

Page 26: ...ate Maximum State Select Yes to calibrate the loaded bin The message Please Wait is displayed Wait until the end of the calibration this process may take a few minutes If calibration is successful the...

Page 27: ...se the control ler displays an error message The stability tolerance is the stability value the controller can tolerate to calibrate the scale By default it is set to 2000 The weight reading may be un...

Page 28: ...in when the weight reading is stable enough Otherwise it will display an error message The stability tolerance is the variation in the weight reading the controller can tolerate to calibrate the bin s...

Page 29: ...cells Sensitivity of Load Cells Enter the sensi tivity of each load cell The sensitivity value is specific to each load cell and is indicated on the load cell s calibration certificate It is also writ...

Page 30: ...rs Re fer to section 6 2 3 to program the timer Worksheets are provided at the end of this manual to write down your time clock relay settings 1 From the main menu select 7 Installation 5 Feed Line Se...

Page 31: ...meter is accessible if the clock schedule is set to Week above Clock Type A time clock can either stop after a specific run time start run or at a pre cise time of the day start stop Choose the desire...

Page 32: ...shes on the display Use the adjustment buttons to set the day 6 Press the right arrow key The hour flashes on the display Use the adjustment buttons to set the hour 7 Proceed in similar fashion to set...

Page 33: ...a menu has more than two items the display scrolls to show additional items Use the right arrow key to select a menu option Use the left arrow key to return to the previous menu Note that if passwords...

Page 34: ...o be completely opened right up against the wall 3 2 Connections 3 2 1 Controller s Main Wiring Refer to the wiring diagram enclosed with this user s manual to connect the controller Use the electrica...

Page 35: ...1 Auger 3 Auger 4 Auger 2 Receiving Hopper 35 TARGET BIN SCALE RF rev 09 TARGET BIN SCALE RF Time Clock Settings Refer to section 6 2 3 to program time clocks and refer to section 5 4 to assign them t...

Page 36: ...ell Ground strap bolted to the bin leg The bin leg should never touch the load cell it must rest on the washers 5 TARGET BIN SCALE RF rev 09 TARGET BIN SCALE RF 2 SYSTEM OVERVIEW 2 1 Description of th...

Page 37: ...al earth terminals which are generally asso ciated with test and measurement circuits These terminals are not for safetyearthingpurposesbutprovide an earth reference point Fuse For Customer Use Enter...

Page 38: ...T BIN SCALE RF rev 09 TARGET BIN SCALE RF TABLE OF CONTENTS 1 INTRODUCTION 4 1 1 Precautions 4 1 2 Symbols of the Manual 4 2 SYSTEM OVERVIEW 5 2 1 Description of the Controller 5 2 2 Typical Applicati...

Page 39: ...M 890 00448 rev 09 REV 08 Target Bin Scale RF User s manual...

Reviews: