NF1200 Manual Edition 1
63
Guidance and manufacturer’s declaration - electromagnetic immunity
The i-PAD is intended for use in the electromagnetic environment specified below. The customer
or the user of the i-PAD should assure that it is used in such an environment.
NOTE 1
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Immunity
Test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
Compliance
level
3 Vrms
10 Vrms
10 V/m
Electromagnetic environment -
guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the i-PAD,
including cables, than the recommended separation
distance calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
where
P
is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and
d
is the recommended separation
distance in metres (m)
Field strengths from fixed RF transmitters, as determined
by an electromagnetic site survey , should be less than
the compliance level in each frequency range .
Interference may occur in the vicinity of equipment
marked with the following symbol:
GG
1.16
d
P
GG
1.2
d
P
1.2
d
P
80MHz to 800MHz
2.3
d
P
800MHz to 2,5GHz
IEC 60601-1 test
level
3 Vrms
150 kHz to 80 MHz
outside ISM bands
10 Vrms
150 kHz to 80 MHz in
ISM bands
10 V/m
80 MHz to 2.5 GHz
a
b
c
d
a
b
c
d
The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6.765 MHz to 6.795 MHz;
13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40,66 MHz to 40.70 MHz.
The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80
MHz to 2.5 GHz are intended to decrease the likelihood that mobile/portable communications equipment could
cause interference if it is inadvertently brought into patient areas. For this reason, an additional factor of 10/3 is
used in calculating the recommended separation distance for transmitters in these frequency ranges.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the i-PAD is used exceeds the
applicable RF compliance level above, the i-PAD should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as or relocating the i-PAD.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.