Do not immerse any part of the i-PAD in fluids.
ื
If the i-PAD has been submerged in water, call CU Medical Systems, Inc. or an authorized
representative for service assistance
ื
Do not allow fluid to enter the case of the device.
ื
Do not spill liquids on the case of the device.
ื
Do not use strong, acetone-based cleaners in cleaning the device.
ื
Do not use abrasive materials in cleaning the unit, especially on the infrared filter of the IrDA
port.
ื
Do not sterilize the i-PAD.
$"65*0/
!
2. Place the AED electrode pads on top of the front panel of the i-PAD. Close the cover down on
the pads.
Cleaning the i-PAD
Clean the i-PAD after each use with a soft cloth which has been moistened with any one of these
approved solutions:
Dilute soap and water
Dilute Chlorine bleach and water mixture (30 ml bleach/liter of water)
Dilute ammonia-based cleaners
Dilute hydrogen peroxide
NF1200 Manual Edition 1
35