10

GG-NH3

Specifications

Input Power:

 +24 VDC, 350 mA

Detection Principle:

 Electrochemical

Detection Method:

 Diffusion

Gases:

 Ammonia (NH3)

Ranges:

0/100 (standard)

0/250

0/300

0/500

0/1000

Custom ranges available. Call for more information

Output Signal:

 Linear 4/20 mA (max input impedance: 

700 Ohms)

Response Time:

T50 = less than 30 seconds

T90 = less than 60 seconds

Accuracy:

 +/- 5% of full-scale, but dependent on 

calibration gas accuracy and time since last calibration

Zero Drift:

 Less than 0.1% of full-scale per month, 

non-cumulative

Span Drift:

 Application dependent, but generally less 

than 3% per month

Linearity:

 +/- 0.5% of full-scale

Repeatability:

 +/- 1% of full scale

SafeCell Viability Check:

 Automatic electrical viability 

test of the electrochemical cell every 40 seconds.  

Failure of the test results in 0.5 mA output after 6 

hours of continued and consecutive test failures.

Power (green) LED:

 Blinks once per second for 60 

seconds during power-up. If supply voltage is too 

low (<10 VDC) or improperly grounded, green LED 

will blink once per second continuously. Stays steady 

on to indicate power. Blinks twice per second in 

calibration mode (4-minute timeout delay)

Status (amber) LED:

 Blinks once per second if RFI 

(radio frequency interference) is detected

Fault (red) LED:

 Stays on steady if supply voltage is 

too low (<10 VDC)

Wiring Connections:

 3-conductor, shielded, 

stranded, ≥ 20 AWG cable (General Cable C2525A or 

equivalent) up to 1500 ft

Terminal Block Plug (Field Wiring):

 26-12 AWG, 

torque 4.5 lbs-in

Enclosure:

 Injection-molded, NEMA 3RX washdown-

duty, polycarbonate sensor housing with hinged lid 

and captive screw. For non-classified areas. 

Optional 18 GA, NEMA 3RX washdown-duty stainless 

steel enclosure with hinged lid and captive screw. For 

non-classified areas

Temperature Range:

 -50˚F to +122˚F (-46˚C to 

+50˚C)

Humidity Range:

 5% to 100% condensing

Dimensions:

 7.7" high x 6.7: wide x 3.8" deep

Weight:

 3.0 lbs

Certification:

 ETL Listed: Conforms to UL 61010-1 

Certified to CSA C22.2 No. 61010-1

Summary of Contents for GG-NH3

Page 1: ...Installation and Operation Manual GG NH3 AMMONIA GAS SENSOR...

Page 2: ...ly in the manner described in this manual If the equipment is used in a manner not specified by Calibration Technologies the protection provided by the equipment may be impaired This equipment should...

Page 3: ...ting the Sensor 4 Installation Guidelines 5 Wiring 6 Operation 7 Start Up 7 Calibration 7 Maintenance 9 Specifications 10 Warranty 11 For technical support contact CTI 920 N Tradewinds Pkwy Columbia M...

Page 4: ...20 mA signal output will drop to 0 5 mA and the fault LED stays on solid This signal should be interpreted as a sensor fault at the control panel and prompt notification to plant personnel to investi...

Page 5: ...wire runs away from mercury vapor lights variable speed drives and radio repeaters to prevent electrical interference Protect sensor from physical damage forklifts etc Mounting screws should be screw...

Page 6: ...ded copper cable Do not pull sensor wiring with AC power cables This can cause electrical interference If cable runs cannot be made without a splice all splice connections should be soldered Ground th...

Page 7: ...replacement unplug cell from the transmitter and adjust signal to 40 mV Span Calibration DO NOT ADJUST THE SPAN POT WITHOUT CERTIFIED CALIBRATION GAS If span adjustment is required use the following...

Page 8: ...ls On steady if supply voltage is less than 10VDC or SafeCell test fails for six hours Status amber LED Blinks once per second if RFI is detected Power green LED On steady to Indicate Power Blinks onc...

Page 9: ...term exposure to ammonia When the SafeCell test fails the fault red LED will blink once per second After six hours of continued and consecutive test failures the 4 20 mA signal output will drop to 0...

Page 10: ...inks once per second for 60 seconds during power up If supply voltage is too low 10 VDC or improperly grounded green LED will blink once per second continuously Stays steady on to indicate power Blink...

Page 11: ...air of the product by any person other than an authorized dealer or contractor or the installation of unapproved parts on the product The obligations set forth in this warranty are conditional on a pr...

Page 12: ...ctiengineering com 866 394 5861 GG NH3 DOC1 3 20190605...

Reviews: