Maintenance / Repair
53
Examination of all functions:
•
temperature measurement
•
displays
•
switching functions automatic and manual filling
•
alarm functions
•
data transmission
•
function of the software
•
relay outputs alarm (I/O box, if required)
•
examination of the set parameters
•
examination of battery voltage, exchange if required
•
examination of the electrical safety
We recommend a conclude a service contract with
Cryotherm GmbH & Co. KG for the yearly maintenance !
Additionally to the yearly maintenance, the following
examinations have to be carried out by the user weekly:
•
visual examination of the vessel and the connecting line
•
check all connections for tightness (visual examinations)
•
check the function of the solenoid valve
•
visual examination of the electrical connecting cables
In addition to the weekly examination, the following
examinations have to be carried out at the BIOSAFE-
CONTROL
®
ß by the user per working day: