Maintenance / Repair
52
Examination and maintenance work at the
BIOSAFE
®
120/220/420 MDß
has to be carried out at regular intervals.
Maintenance and repair work may only be carried
out by trained and instructed skilled personal.
Visual examinations and operational checks can be
carried out by instructed users (operators).
Service Hotline
02741 - 95 85 75
Once a year, a maintenance has to be carried out.
This maintenance may only be carried out by:
•
Cryotherm service personnel
•
Skilled personnel specially trained for this medical
device by the manufacturer
CHRONOS
®
:
•
visual examination (outer condition, stability)
•
examination for undue frost formation
•
examination of the castors
•
examine the cover for damage (handles, insulation, outer
condition)
•
examine the pressure in the gas pressure spring of the
cover opening mechanism
•
visual examination (outer condition) of the LIN line
•
clean the solenoid valve, check the function
•
clean the LIN filter
BIOSAFE-CONTROL
®
ß:
Visual examination
•
examination of outer condition
•
fastening
•
plug-in connections
•
cables
•
leadthroughs
9
Maintenance / Repair
9.1 Scope of Maintenance
Summary of Contents for BIOSAFE 120
Page 1: ...Operating Manual BIOSAFE 120 220 420 MD Service Hotline 49 0 2741 95 85 75...
Page 2: ......
Page 31: ...Safety 27 5 1 Safety Safety advices Handling with cryogenic liquefied gases Source IGV Germany...
Page 32: ...Safety 28...
Page 33: ...Safety 29...
Page 51: ...Accessories 47 660 mm 95 mm 165 mm CHRONOS 220 Towers 77031511 95 x 165 mm...