Cryo-con Model 24C
Specifications, Features and Functions
CalGen Calibration Curve Generator
The CalGen feature generates new calibration curves for Silicon diode, thermocouple
or Platinum sensors. This provides a method for obtaining higher accuracy
temperature measurements without expensive sensor calibrations.
Curves can be generated from any user-selected curve and are written to a specified
internal user calibration curve area.
The CalGen function may be performed in the instrument by using the front panel.
Alternatively, the feature is also implemented in the Cryo-con utilities software.
Input Channel Statistics
Input temperature statistics are continuously maintained on each input channel. This
data may be viewed in real time on the Input Channel menu, or accessed via any of
the remote I/O ports.
Statistics are:
Minimum Temperature.
Maximum Temperature.
Temperature Variance.
Slope and Offset of the best-fit straight line to temperature history.
Accumulation Time
The temperature history may be cleared using a reset command provided.
Electrical Isolation and Input Protection
The input channel measurement circuitry is electrically isolated from other internal
circuits. However, the common mode voltage between an input sensor connection
and the instrument's ground should not exceed
40V.
Sensor inputs and outputs are provided with protection circuits. The differential
voltage between sensor inputs should not exceed
15V.